Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Class |
|---|---|
| Shanghai 6th People's Hospital | OTHER |
| Shanghai East Hospital | OTHER |
This study is to support the regulatory application of OneTouch Select® Plus test strips in China. The blood glucose test strip is regulated as an In Vitro Diagnostic (IVD) product and China Food and Drug Administration (CFDA) regulation of IVDs should be complied with. The result of this study will be used to support regulatory filing in China.
Investigational Product:
OneTouch Select® Plus test strips (test with OneTouch Select® Plus glucose meter)
Objectives:
Comparator(control) Product:
YSI 2300 glucose analyser
Acceptance Criteria:
User Performance of the OneTouch Select® Plus Blood Glucose Monitoring system(described as OneTouch Select Plus below) is met if the lay user and HCP results obtained with OneTouch Select Plus meet the requirement stated in section 7.4.1 of ISO () 15197:2003(E). Further to meet the requirements of ISO 15197:2013(E) , the lay user results obtained with OneTouch Select Plus should meet the requirements stated in section 8.2 of ISO 15197:2013(E). The effectiveness of instructions for use is met , if for each item on the Instructions for Use Questionnaire, the 90% lower confidence limit (90% LCL) for the correct response rate is equal to or greater than 70%.
Study Design and subject population:
This study is an open, non-randomized clinical evaluation and will be conducted at a minimum of 2 sites in China. At least 240 evaluable diabetic subjects who meet the enrolment criteria will participate in this evaluation.
Study procedures:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| OneTouch Select® Plus Glucose Test Strips | Device |
| Measure | Description | Time Frame |
|---|---|---|
| The accuracy of glucose results obtained from OneTouch Select Plus BGMS | User Performance of the OneTouch Select® Plus Blood Glucose Monitoring system(described as OneTouch Select Plus below) is met if the lay user and HCP results obtained with OneTouch Select Plus meet the requirement stated in section 7.4.1 of ISO 15197:2003(E).Further to meet the requirements of ISO 15197:2013(E) , the lay user results obtained with OneTouch Select Plus should meet the requirements stated in section 8.2 of ISO 15197:2013(E). | instantly after Glucose testing |
| The correct response rate of Instructions for Use Questionnaire | The effectiveness of instructions for use is met , if for each item on the Instructions for Use Questionnaire, the 90% lower confidence limit (90% LCL) for the correct response rate is equal to or greater than70% | instantly after Glucose testing |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided