This study scope is to validate the accuracy and usability of the Soberlink Cellular Device in comparison to a predicate device, BACtrack S80 Pro.
The purpose of this accuracy and usability study is to validate the performance of Soberlink Cellular Device in the hands of the intended (lay) users - untrained study participants who have consumed alcohol and use the Soberlink Cellular Device to measure their blood alcohol content (BAC) in comparison to a predicate device.
The purpose of this document is to define how the human factors environment impacts the actual user interface design as well as how usability engineering evaluates user interface via usability objectives and usability tests. This accuracy and usability study will be performed on Soberlink Cellular Device in order to comply with FDA's Draft Guidance for Industry and Food and Drug Administration Staff, titled: Applying Human Factors and Usability Engineering to Optimize Medical Device Design, IEC 60601-1-6 (collateral standard for usability engineering of medical electrical equipment), and IEC 62366 (application of usability engineering to medical devices).
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Soberlink Cellular Device | Active Comparator | BAC reading with Soberlink Cellular Device |
|
| BACtrack S80 Pro | Active Comparator | BAC reading with BACtrack S80 Pro |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Soberlink Cellular Device | Device | BAC breath analyzer |
| |
| Measure | Description | Time Frame |
|---|---|---|
| Validate the accuracy of the Soberlink Cellular Device in comparison to a predicate device, BACtrack S80 Pro and assessed by linear regression. | Performance validation will occur by evaluating the accuracy of the Soberlink Cellular device in the hand of intended users - untrained study participants who have consumed alcohol and use the device to measure their blood alcohol content (BAC). 40 adults will be recruited who have recently consumed alcohol. The only assistance to be provided is the labeling. No additional coaching, either written or verbal, will be provided to study participants. Participants should obtain a BAC reading with the Soberlink Cellular device, and a second BAC reading with a comparator device. The second measurement will be taken within 5 minutes of the first. A linear regression line for the comparison between results of the Soberlink Cellular device and a comparator will be made. | Day 1 |
| Measure | Description | Time Frame |
|---|---|---|
| Validate the usability of the Soberlink Cellular Device in comparison to a predicate device, BACtrack S80 Pro and assessed by Questionnaire. | Performance validation will occur by evaluating the usability of the Soberlink Cellular device in the hand of intended users - untrained study participants who have consumed alcohol and used the device to measure their blood alcohol content (BAC). 40 adults who have recently consumed alcohol will be asked to evaluate the Soberlink Cellular device. The only assistance provided is the labeling. No additional coaching, either written or verbal, will be provided to study participants. Participants should obtain a BAC reading with the Soberlink Cellular device, and a second BAC reading with a comparator device. The second measurement should be taken within 5 minutes of the first. Following the second measurement a questionnaire will be administered to study participants to evaluate the ease of use of the Soberlink Cellular device. |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Shannon Hanrahan, PhD | Private Practice | Principal Investigator |
Not provided
Participant will be aware of their own BAC result as the result is displayed on the LED screen of the device
Not provided
Not provided
Not provided
Not provided
| Type | Date | Date Unknown |
|---|---|---|
| Release | Aug 22, 2017 | |
| Reset | Sep 19, 2017 | |
| Release | Mar 14, 2018 | |
| Unrelease | Yes | |
| Release | Apr 16, 2018 | |
| Reset | May 15, 2018 |
Not provided
Not provided
| Release Date | Unrelease Date | Unrelease Date Unknown | Reset Date | MCP Release Number |
|---|---|---|---|---|
| Aug 22, 2017 | Sep 19, 2017 | |||
| Mar 14, 2018 |
| ID | Term |
|---|---|
| D019966 | Substance-Related Disorders |
| ID | Term |
|---|---|
| D064419 | Chemically-Induced Disorders |
| D001523 | Mental Disorders |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| BACtrack S80 Pro |
| Device |
BAC breath analyzer |
|
| Day 1 |
| Yes |
| Apr 16, 2018 | May 15, 2018 |