Background During ventilatory assistance, optimization of settings is critical to allow a personalized support and avoid over- or under-assistance. But little data are available in clinical practice to guide the adjustment of the support. In adults, esophageal pressure (PES) has been shown to be a reliable surrogate of pleural pressure (PPL) and clinical studies suggest that PES may be useful to guide the management of mechanical ventilation. In children, the PES measurement could have similar potential benefits, but beforehand the reliability of PES to estimate PPL needs to be assessed.
Objective The primary objective of this study is to validate the reliability of PES directly monitored using a miniature catheter tip pressure transducer (Gaeltec® system) to estimate PPL, when compared to a gold standard, i.e the direct PPL measurement in situ.
Method This is a prospective single center study. Children <18 years old, hospitalized in the pediatric intensive care unit, requiring invasive ventilation and with at least one chest tube will be included.
Protocol A pressure transducer will be connected to the existing chest-tube and PES (measured by Gaeltec® and feeding tube), PPL, PAW, respiratory volume and flow will be simultaneously recorded.
Expected results We expect that the PES-based methods will provide an accurate estimation of PPL. Once this tool validated, PES could be helpful to optimize mechanical ventilation in children, and further interventional trials would be warranted to evaluate if its use could allow a reduction of the ventilation support duration.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Patients | Experimental |
|
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| esophageal and pleural pressure measurement | Other | Esophageal pressure will be monitored by a dedicated catheter and pleural pressure will be measured into the chest tube already in place |
| Measure | Description | Time Frame |
|---|---|---|
| End-expiratory esophageal | End-expiratory PES and PL, calculated as the mean value of 10 consecutive breaths, during a stable period with no intervention or artifacts; | Up to 2 hours from the beginning of the study |
| Pleural pressure | PPL, directly measured in the existing chest tubes, and calculated as the mean value of 10 consecutive breaths, during a stable period with no intervention or artifacts. | Up to 2 hours from the beginning of the study |
| transpulmonary pressures | Up to 2 hours from the beginning of the study |
| Measure | Description | Time Frame |
|---|---|---|
| The Elastance-derived end-inspiratory transpulmonary pressure | The Elastance-derived end-inspiratory transpulmonary pressure, calculated as: PAW at end-inspiration x EL/ERS (EI and ERS being the respective elastances of the lung and the respiratory system, estimated on the esophageal and airway Pressure-Volume curves); | Up to 2 hours from the beginning of the study |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| St. Justine's Hospital | Montreal | Quebec | H3T 1C5 | Canada |