The trial will investigate the accuracy and performance of a new sensor for continuous glucose monitoring (CGM), the (new generation) Enlite, CE-marked in 2013, in conjunction with the new Medtronic 640G system components.
To evaluate the accuracy and performance of the new generation Enlite sensor. This is a postmarket, interventional, open-label, non-randomized, single center clinical research trial.
The trial will adopt a prospective single-sample design using the Yellow Spring Instruments (YSI) plasma glucose reference sampling procedure in the clinic and self-monitoring of blood glucose (SMBG) plasma glucose reference sampling procedure at home.
The trial will investigate the accuracy and performance (i.e. the life-time) of a new sensor for CGM, the (new generation) Enlite glucose sensor, CE-marked in 2013, in conjunction with the new Medtronic 640G system components. Accuracy and life-time of the sensor are important parameters contributing to the safety of patients and patient acceptance of this still fairly new technology.
The current trial investigates the accuracy of the new sensor in conjunction with the new 640G system components both in the clinic and at home. This enables us to assess the sensor accuracy in a well-controlled environment using a highly accurate laboratory glucose analyzer as reference as well as sensor accuracy during real-life conditions using a blood glucose meter as reference analyzer.
The data gathered in this trial will help to understand the effect of technological advancements made and can guide future development of glucose sensor for continuous glucose monitoring.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Open-label, single-sample design | Other | The trial has an open label design with a glucose sensor intervention. Only 1 sensor system is used in the trial (no comparator). As all subjects will receive identical devices and undergo the same experimental procedures, no randomization will be performed. Furthermore, neither subjects nor clinical staff will be blinded to the sensor readings, as knowing the sensor glucose readings will not affect the accuracy endpoint. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Glucose sensor (Enlite) | Device | This is a postmarket, interventional, open-label, non-randomized, single center clinical research trial. The trial will adopt a prospective single-sample design using the Yellow Spring Instruments (YSI) plasma glucose reference sampling procedure in the clinic and selfmonitoring of blood glucose (SMBG) plasma glucose reference sampling procedure at home. The trial will consist of 6 visits to the investigational site: a visit for informed consent procedures (Visit 0), a screening visit (Visit 1), 3 study visits (Visit 2-4) and a final examination (Visit 5). |
| Measure | Description | Time Frame |
|---|---|---|
| Sensor Accuracy | Sensor values were compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing day (day 3). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). | 4 months |
| Consensus Error Grid Analysis of Paired Sensor and Reference Plasma Glucose Values | All analysis performed using the Consensus Error Grid (or Parkes error grid) comparing the paired sensor and YSI reference glucose values. Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action." Zone B as "altered clinical action with little or no effect on clinical outcome.". Ideal situation is 100% in Zone A + B. | 4 months |
| Sensor Survival | sensor survival in hours | 4 months |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
All Subjects
All Subjects
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | All Subjects | All subjects will receive identical devices and undergo the same experimental procedures. |
| Title | Milestones | Reasons Not Completed | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | All Subjects | All Subjects |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Continuous | Mean |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Sensor Accuracy | Sensor values were compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing day (day 3). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). | Posted | Mean | Standard Deviation | percentage | 4 months |
|
|
Not provided
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | All Subjects | 0 | 24 |
Not provided
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| Abdominal Pain | Gastrointestinal disorders | Systematic Assessment |
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Odile O'Sullivan, Clinical Trial Manager | Medtronic Minimed | 8184763319 | odile.osullivan@medtronic.com |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
| Years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Region of Enrollment | Count of Participants | Participants |
|
|
|
| Primary | Consensus Error Grid Analysis of Paired Sensor and Reference Plasma Glucose Values | All analysis performed using the Consensus Error Grid (or Parkes error grid) comparing the paired sensor and YSI reference glucose values. Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action." Zone B as "altered clinical action with little or no effect on clinical outcome.". Ideal situation is 100% in Zone A + B. | Posted | Number | percentage | 4 months |
|
|
|
| Primary | Sensor Survival | sensor survival in hours | Posted | Mean | 95% Confidence Interval | hours | 4 months |
|
|
|
| 0 |
| 24 |
| 5 |
| 24 |
| itching at sensor insertion | General disorders | Systematic Assessment |
|
| hematoma at insertion | General disorders | Systematic Assessment |
|
| erythema at insertion site | General disorders | Systematic Assessment |
|
| mild erythema | Skin and subcutaneous tissue disorders | Systematic Assessment |
|