Continuous haemoglobin determination could improve the outcomes of several patients in risk of bleeding during surgery. An in vivo calibration has been introduced to the device after our team found its need. The uncalilbrated validation of the Masimo Radical 7 continuous haemoglobin monitor has been made in various papers, although it has been compared to point of care devices, thus introducing an error as such point of care devices although validated for haemoglobin determination, are not as accurate as the gold standard. The validation of the gold standard calibrated device has not been described yet.
When compared to the accepted gold standard (cyanmethemoglobin method, Coulter), accuracy and precision of the continuous haemoglobin monitor could be good enough to be used interchangeably with the gold standard.
Continuous haemoglobin determination could improve the outcomes of several patients in risk of bleeding during surgery.
The in vivo calibration has not been adjusted to the gold standard haemoglobin determination in the literature yet.
Two blood samples will be withdrawn at the same time from an arterial line, one for the central laboratory and the other one for the POC device. At the same exact time the SpHb measurement will be recorded. At least eight samples are expected per patient. A maximum of 20 samples will be withdrawn per patient.
To compare bias, precision and limits of agreement among SpHb, LabHb and COoxHb the Bland Altman method will be used with multiple observations per subject when the true value varies. An error grid analysis will be made to know how it could affect to clinical decision making.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Masimo Radical 7 | Device | Simultaneous determination of haemoglobin with the continuous haemoglobin monitor, the central lanboratory coulter and a POC: |
| Measure | Description | Time Frame |
|---|---|---|
| Central coulter bias and limits of agreement | Bias and limits of agreement of SpHb compared to the central laboratory coulter | 3 months |
| Central coulter bias and limits of agreement after in vivo calibration | Bias and limits of agreement of SpHb compared to the central laboratory coulter after in vivo Radical 7 calibration | 3 months |
| Measure | Description | Time Frame |
|---|---|---|
| POC Cooximeter bias and limits of agreement | Bias and limits of agreement of SpHb compared to the POC cooximeter | 3 months |
| POC Cooximeter bias and limits of agreement | Bias and limits of agreement compared to the POC cooximeter after in vivo Radical 7 calibration |
| Measure | Description | Time Frame |
|---|---|---|
| Comparison of bias and limits of agreement | Comparison of bias and limits of agreement among central laboratory, POC cooximeter and SpHb continuous monitor. | 3 months |
| Clinical decision making | Error grid analysis plot |
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Adult patients scheduled to any surgery of at least three hours of duration.
| Name | Role | Phone | Extension | |
|---|---|---|---|---|
| Juan Soliveres, MD, PhD | Contact | +34670758871 |
| Name | Affiliation | Role |
|---|---|---|
| Juan Soliveres, MD, PhD | University Hospital Doctor Peset | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| University Hospital Doctor Peset | Recruiting | Valencia | 46023 | Spain |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| Background | Soliveres J, Balaguer J, Estruch M, Sánchez A, Sánchez J. Validation of Continuous and Noninvasive Hemoglobin Monitoring from Pulse CO-Oximetry during Surgery. 2010 Annual Meeting of the American Society Anesthesiologists |
| Label | URL |
|---|---|
| Citation link | View source |
Not provided
Not provided
Not provided
Not provided
Not provided
| 3 months |
| POC Cooximeter Coulter interchangeability | Bias and limits of agreement of POC Cooximeter Hb determination compared to the central laboratory coulter | 3 months |
| 3 months |