The purpose of this study was to determine if untrained subjects with diabetes and without diabetes could operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Persons With Diabetes | Experimental | Untrained Subjects WITH Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| ONYX NEXT BGMS | Device | Untrained Persons WITH Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX NEXT BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. Also, study staff tested venous blood from subjects WITH Diabetes and BG results were compared to reference method results obtained from the subjects' venous plasma. |
| Measure | Description | Time Frame |
|---|---|---|
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (<100 mg/dL YSI capillary plasma) and +/- 15% (>=100 mg/dL YSI capillary plasma). | 1 hour |
| Measure | Description | Time Frame |
|---|---|---|
| Percent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Untrained subjects WITH Diabetes (332) self-tested Alternate Site (AST) palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (<100 mg/dL YSI capillary plasma) and +/- 15% (>=100 mg/dL YSI capillary plasma). |
Not provided
Inclusion Criteria:
Exclusion Criteria:
General enrollment guidelines:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Timothy Bailey, MD, FACP | AMCR Institute | Principal Investigator |
| Mark Christiansen, MD | Diablo Clinical Research | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| AMCR Institute | Escondido | California | 92026 | United States | ||
| Diablo Clinical Research |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Persons With and Without Diabetes | Untrained Subjects WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). The ONYX NEXT BGMS: Untrained Persons WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX NEXT BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. |
| Title | Milestones | Reasons Not Completed | ||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Persons With and Without Diabetes | Untrained Subjects WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). The ONYX NEXT BGMS: Untrained Persons WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX NEXT BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Continuous | Mean |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (<100 mg/dL YSI capillary plasma) and +/- 15% (>=100 mg/dL YSI capillary plasma). | 329 (332 with diabetes-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG result after three attempts. | Posted | Number | Percent of Results within 15mg/dL/15% | 1 hour |
|
Not provided
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Persons With and Without Diabetes | Untrained Subjects WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). The ONYX NEXT BGMS: Untrained Persons WITH and WITHOUT Diabetes Use the ONYX NEXT BGMS (Blood Glucose Monitoring System). Subjects tested capillary fingerstick and palm blood (and study staff tested subject fingerstick blood) using the ONYX NEXT BGMS. All BG results were compared to reference method results obtained with subject capillary plasma. |
Not provided
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| Hypoglycemia | Endocrine disorders | Non-systematic Assessment | Subjects with hypoglycemia (defined by site as BG level < 70 mg/dL) had diabetes and hypoglycemia was resolved at the visit. |
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Jane Wallace, Deputy Director for Global Clinical Affairs | Ascensia Diabetes Care | 574-257-3063 | jane.wallace@ascensia.com |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
| 1 hour |
| Percent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff | Study staff tested subject (332 WITH diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (<100 mg/dL YSI capillary plasma) and +/- 15% (>=100 mg/dL YSI capillary plasma). | 1 hour |
| Percent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff | Study staff tested venous blood of 332 subjects WITH diabetes using an Investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (<100 mg/dL YSI venous plasma) and +/- 15% (>=100 mg/dL YSI venous plasma). | 1 hour |
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | Untrained subjects WITH Diabetes (332) and WITHOUT Diabetes (43) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range. | 1 hour |
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested Fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range. | 1 hour |
| Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range. | 1 hour |
| Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range. | 1 hour |
| Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff | Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory method across the entire tested YSI glucose range. | 1 hour |
| Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff | Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory method across the entire tested YSI glucose range. | 1 hour |
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method | Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5 mg/dL (<100 mg/dL YSI capillary plasma) and +/- 12.5% (>=100 mg/dL YSI capillary plasma). | 1 hour |
| Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Staff obtained responses from persons WITH Diabetes (332) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'. | 1 hour |
| Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Staff obtained responses from persons WITH and WITHOUT Diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'. | 1 hour |
| Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Staff obtained responses from persons WITH Diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'Choose Not to Answer'. | 1 hour |
| Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method | Untrained subject WITH and WITHOUT diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (< 75 mg/dL YSI capillary plasma) and +/- 15% (>= 75 mg/dL YSI capillary plasma). | 1 hour |
| Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method | Untrained subject WITH and WITHOUT diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (< 75 mg/dL YSI capillary plasma) and +/- 15% (>= 75 mg/dL YSI capillary plasma). | 1 hour |
| Walnut Creek |
| California |
| 94598 |
| United States |
| years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Race/Ethnicity, Customized | Number | Number of responses |
|
| Region of Enrollment | Number | participants |
|
|
|
| Secondary | Percent of Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Untrained subjects WITH Diabetes (332) self-tested Alternate Site (AST) palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (<100 mg/dL YSI capillary plasma) and +/- 15% (>=100 mg/dL YSI capillary plasma). | 318 (332 with diabetes -14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol. | Posted | Number | Percent of Results within 15mg/dL/15% | 1 hour |
|
|
|
| Secondary | Percent of Subject Fingerstick Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff | Study staff tested subject (332 WITH diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (<100 mg/dL YSI capillary plasma) and +/- 15% (>=100 mg/dL YSI capillary plasma). | 332 Blood glucose results from subjects with diabetes were analyzed. | Posted | Number | Percent of Results within 15mg/dL/15% | 1 hour |
|
|
|
| Secondary | Percent of Venous Blood Glucose (BG) Results (From Subjects WITH Diabetes) Within +/- 15mg/dL (<100mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Tested by Study Staff | Study staff tested venous blood of 332 subjects WITH diabetes using an Investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results were compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (<100 mg/dL YSI venous plasma) and +/- 15% (>=100 mg/dL YSI venous plasma). | 330 (332-2) Blood glucose results were analyzed. Two (2) subjects had unsuccessful venipuncture attempts, so no venous results were obtained for them. | Posted | Number | Percent of Results within 15mg/dL/15% | 1 hour |
|
|
|
| Secondary | Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | Untrained subjects WITH Diabetes (332) and WITHOUT Diabetes (43) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range. | 372 (375-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG results after three attempts. | Posted | Number | Percent of Results within 15% | 1 hour |
|
|
|
| Secondary | Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested Fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range. | 372 (375-3) Blood glucose results were analyzed. Three subjects did not contain meter BG results after three attempts. | Posted | Number | Percent of Results within 20% | 1 hour |
|
|
|
| Secondary | Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method Across the Tested Glucose Range | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory reference method across the entire tested YSI glucose range. | 361 (375-14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol. | Posted | Number | Percent of Results within 15% | 1 hour |
|
|
|
| Secondary | Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method Across the Tested Glucose Range | Untrained subjects WITH diabetes (332) and WITHOUT diabetes (43) self-tested AST palm blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory reference method across the entire tested YSI glucose range. | 361 (375-14) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. Five (5) subjects had low blood sugar; their palm results were not evaluable per protocol. | Posted | Number | Percent of Results within 20% | 1 hour |
|
|
|
| Secondary | Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15% of Laboratory Glucose Method When Tested By Study Staff | Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15% of the laboratory method across the entire tested YSI glucose range. | 375 Blood glucose results were analyzed. | Posted | Number | Percent of Results within 15% | 1 hour |
|
|
|
| Secondary | Percent of Subject Fingerstick Blood Glucose (BG) Results Within +/- 20% of Laboratory Glucose Method When Tested By Study Staff | Study staff tested subject (332 WITH and 43 WITHOUT diabetes) fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 20% of the laboratory method across the entire tested YSI glucose range. | 375 Blood glucose results were analyzed. | Posted | Number | Percent of Results within 20% | 1 hour |
|
|
|
| Secondary | Percent of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 12.5mg/dL (<100mg/dL) and Within +/- 12.5% (>=100 mg/dL) of Laboratory Glucose Method | Untrained subjects WITH Diabetes (332) self-tested fingerstick blood using an Investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 12.5 mg/dL (<100 mg/dL YSI capillary plasma) and +/- 12.5% (>=100 mg/dL YSI capillary plasma). | 329 (332-3) Blood glucose results were analyzed. Three subjects did not obtain meter BG result after three attempts. | Posted | Number | Percent of Results within12.5mg/dL/12.5% | 1 hour |
|
|
|
| Secondary | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Staff obtained responses from persons WITH Diabetes (332) using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'. | Posted | Number | Percent of Subjects who responded | 1 hour |
|
|
|
| Secondary | Percent of Responses From Persons WITH and WITHOUT Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding BGMS | Staff obtained responses from persons WITH and WITHOUT Diabetes using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree'. | Posted | Number | Percent of Subjects who responded | 1 hour |
|
|
|
| Secondary | Percent of Responses From Persons WITH Diabetes That Either 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements Regarding Views/Behaviors Related to Self-Monitoring Blood Glucose | Staff obtained responses from persons WITH Diabetes using short questionnaires to provide feedback on views and behaviors related to managing their diabetes. Subjects could respond 'Strongly Agree' or 'Agree' or are 'Neutral' or 'Disagree' or 'Strongly Disagree' or 'Choose Not to Answer'. | Posted | Number | Percent of Subjects who responded | 1 hour |
|
|
|
| Secondary | Percent of Self-Test Fingerstick Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method | Untrained subject WITH and WITHOUT diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (< 75 mg/dL YSI capillary plasma) and +/- 15% (>= 75 mg/dL YSI capillary plasma). | 372 (375 - 3) Blood glucose results were analyzed. Three subjects did not obtain meter BG results after three attempts. | Posted | Number | Percent of Results within 15mg/dL/15% | 1 hour |
|
|
|
| Secondary | Percent of Self-Test Alternate Site Palm Blood Glucose (BG) Results (From Subjects WITH and WITHOUT Diabetes) Within +/- 15 mg/dL (<75 mg/dL) and Within +/- 15% (>= 75 mg/dL) of Laboratory Glucose Method | Untrained subject WITH and WITHOUT diabetes self-tested Alternate Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the percent of BGMS results within +/- 15 mg/dL (< 75 mg/dL YSI capillary plasma) and +/- 15% (>= 75 mg/dL YSI capillary plasma). | 366 (375-9) Blood glucose results were analyzed. Nine (9) subjects with low blood sugar did not attempt palm testing per protocol. | Posted | Number | Percent of Results within 15mg/dL/15% | 1 hour |
|
|
|
| 0 |
| 375 |
| 8 |
| 375 |
|
Not provided
| Title | Measurements |
|---|---|
|
| I like the overall meter design. |
|
| I find the meter easy to use. |
|
| Instructions (User Guide) were easy to understand. |
|
| Instructions clearly explain how to run a test. |
|
| Instructions clearly explain meter error messages. |
|
| Title | Measurements |
|---|---|
|
| I like the overall meter design. |
|
| I find the meter easy to use. |
|
| Instructions (User Guide) were easy to understand. |
|
| Instructions clearly explain how to run a test. |
|
| Instructions clearly explain meter error messages. |
|
| Title | Measurements |
|---|---|
|
| Acc:important to help prevent low blood sugar |
|
| Acc:important to understand food/exercise effects. |
|
| Acc:help w using results to better control diabete |
|
| Acc:important to achieve greater peace of mind. |
|
| I prefer the meter I just used to my regular meter |
|
| I use my meter - my care provider gave it to me. |
|
| I use my current meter -my insurance covers strips |
|
| I use my current meter-I think it is most accurate |
|