The objective of this study is to demonstrate that the OptiScanner® is safe and can provide accurate blood glucose levels in critically ill subjects.
Accuracy Hypothesis:
The assessment of blood glucose level that results from the OptiScanner is comparable to the YSI 2300 STAT Plus™ Glucose and Lactate Analyzer ("YSI Analyzer"; YSI Life Sciences, Yellow Springs, OH).
Safety Hypothesis:
The OptiScanner has an acceptable risk/benefit profile for a system that can provide repeated automated blood glucose levels for critically ill subjects.
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Optiscanner values vs YSI | Experimental | Matched samples up to 12 times per 24 hour period |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| The OptiScanner | Device | treatment of blood glucose per standard of care |
|
| Measure | Description | Time Frame |
|---|---|---|
| Primary Accuracy Endpoint | The mean absolute relative difference (MARD) in glucose level between the OptiScanner and the YSI Analyzer (average of 2 measures). To demonstrate accuracy the MARD must be less than or equal to 10%. | Accurancy up to 72 hours of monitoring per subject |
| Measure | Description | Time Frame |
|---|---|---|
| Secondary Accuracy Endpoint | The "population CV" will be computed as the standard deviation of the difference between the glucose level reported by the OptiScanner and the glucose level reported by the YSI Analyzer divided by the mean of the YSI Analyzer values. Repeated measures analyses will be used to account for possible correlation of measurements within patients. The population CV therefore gives an estimate of the extent of the effect of outliers on the accuracy estimation. If the "population CV" is less than 13%, the outcome of the secondary accuracy analysis will be considered a success. |
| Measure | Description | Time Frame |
|---|---|---|
| Number of Adverse events | up to 72 hours of subject monitoring |
Inclusion Criteria:
Exclusion Criteria:
Pregnant or nursing
In the Investigator's opinion the subject cannot safely tolerate the amount of saline required to be given to the subject (up to 360 mL per day)
Hematocrit less than 15% or greater than 60%
Subjects that require placement of an additional vascular access line will be excluded if they do not have a suitable access site free from any of the following conditions:
Any medical condition that, in the Investigator's opinion, would warrant exclusion from the study or prevent the subject from completing the study, including but not limited to a high risk of complications associated with vascular access; or
Participation in any other investigational drug or device study in the last 30 days and/or while enrolled in this study.
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Role | Phone | Extension | |
|---|---|---|---|---|
| Dana Deyette, RN | Contact | 510-342-5836 | ddeyette@optiscancorp.com |
| Name | Affiliation | Role |
|---|---|---|
| Grant V Bochicchio, MD, MPH | Washington University School of Medicine | Principal Investigator |
| Stanley A Nasraway, MD | Tufts Medical Center | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Tufts Medical Center | Recruiting | Boston | Massachusetts | 02111 | United States |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 27394384 | Derived | Nohra E, Buckman S, Bochicchio K, Chamieh J, Reese S, Merrill C, Schuerer D, Bochicchio GV. Results of a near continuous glucose monitoring technology in surgical intensive care and trauma. Contemp Clin Trials. 2016 Sep;50:1-4. doi: 10.1016/j.cct.2016.07.007. Epub 2016 Jul 6. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| YSI 2300 STAT Plus™ | Device |
|
| Accuracy up to 72 hours of subject monitoring |
| Anthony Furnary, MD |
| Providence St. Vincents Medical Center |
| Principal Investigator |
| Linda Moore, MD | Hermann Memorial Hospital | Principal Investigator |
| Washington University School of Medicine | Recruiting | St Louis | Missouri | 63108 | United States |
|
| Providence Heart and Vascular Institute | Recruiting | Portland | Oregon | 64111 | United States |
|
| Hermann Memorial Hospital | Recruiting | Houston | Texas | 77030 | United States |
|