This evaluation will evaluate a new system from YOFiMeter, known as YOFiLife Blood Glucose Monitoring System with YOFiMeter glucometer. YOFiLife includes a web- based data system known as the Web Portal and smart phone apps for the iPhone and Android. The new YOFiMeter is similar to other glucometers with the exception of built in test cassettes, lancet cassettes with enhanced communications and data reviewing options.
This evaluation will also collect data from a YSI 2300 Stat Plus Glucose Analyzer reference glucometer. The YSI 2300 Stat Plus Glucose Analyzer will be used as the "gold" standard with all results compared to the YSI results.
This is a single evaluation to assess both the Yofimeter system accuracy in the hands of the intended users, as well as other aspects to support lay use, such as labeling assessment and usability. The company is presenting diabetics who monitor their blood glucose with new technological innovations aimed at improving their daily lives (e.g., a unique blood glucose monitor, Web Portal applications and "smart phone" applications).
Each subject selected for the comparison evaluation will perform their own fingerstick and test on their subject device following device instructions. A trained technician will perform a fingerstick on the subject immediately after and perform the test on the same device. Separate results will be recorded by subject and technician. A trained technician will take another blood sample, within as short of a time as practicable, for a glucose assay on the YSI 2300 Stat Plus Glucose Analyzer.
Not provided
Not provided
Not provided
Not provided
Not provided
| Measure | Description | Time Frame |
|---|---|---|
| Accuracy of the Yofimeter as compared to the YSI 2300 Stat Plus Glucose Analyzer | The patient- and provider- measures of blood glucose will be compared to the YSI Analyzer measures using the standard statistical methods for method-validation analysis: Bland-Altman agreement and Passing-Bablok regression. | Within the hour |
| Measure | Description | Time Frame |
|---|---|---|
| Evaluate the usability of instructional materials. | Evaluate the usability of the instructional materials in assisting a new user to set up and use the YOFiLife Blood Glucose Meter and the data monitoring systems, on the Web Portal, and mobile iPhone and Android apps. | Within 2 hours |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Individuals over 18 years of age with diabetes who currently use blood glucose monitors.
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Timothy Bailey, MD | AMCR Institute | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Profil Institute for Clinical Research | Chula Vista | California | 91911 | United States | ||
| AMCR Institute |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
| Escondido |
| California |
| 92026 |
| United States |
| Tri City Endocrinology | Oceanside | California | 92056 | United States |