The current endoscopic methods for diagnosing and monitoring treatment response in Eosinophilic Esophagitis (EoE) are costly, inconvenient, and risky. Novel diagnostic methods are needed, and the minimally-invasive Cytosponge holds great promise. It has been shown to be safe and accurate in Barrett's esophagus, it has the advantage (over the string test) of obtaining a true tissue sample, and our preliminary data supports its further study in EoE. The proposed prospective cohort study, conducted by experts in esophageal diseases and EoE, will assess the accuracy of Cytosponge compared to endoscopy and biopsy in EoE, and determine the safety and acceptability of this technique. Use of the Cytosponge would fundamentally change the paradigm for clinical management of EoE by allowing collection of non-endoscopic esophageal biopsies, thus minimizing the need for invasive testing. It would also facilitate future genetic, mechanistic, and pathogenesis research in EoE.
Study design overview (all Aims) This will be a prospective cohort study, with patient enrollment conducted at UNC and Mayo Clinic with sample analysis performed by the University of Cambridge. In Aim 1, patients with EoE will be enrolled, tissue will be obtained from both the Cytosponge and endoscopy, and the methods will be compared for a single time point to determine accuracy of Cytosponge for quantifying esophageal eosinophil counts. For all patients, safety will be monitored and subjects will complete a survey about the acceptability of Cytosponge (Aim 2).
Cytosponge protocol:
After the study has been explained and a patient provides informed consent, the Cytosponge will be administered prior to endoscopy by trained research staff under physician supervision. If subjects opt to receive a local anesthetic, then they will be provided with a 2% lidocaine gargle prior to administration of the Cytosponge. The Cytosponge will be administered according to it's instructions for use. After retrieval, the string is cut and the sponge (which contains the tissue specimen) is placed in a container, immersed in fixative, and stored in a refrigerator at 4°C. The fixative is then spun in a centrifuge, and the pelleted cells are embedded in a paraffin block using standard techniques.
Upper endoscopy and biopsy:
After the Cytosponge has been removed, the patient will undergo standard of care (routine care) upper endoscopy and biopsy, as clinically indicated. During this exam, research staff will record all endoscopic features of EoE, including rings, furrows, white plaques, decreased vascularity, and strictures. The severity of the endoscopy findings will be measured using the recently validated endoscopic reference score (EREFS) scoring system. Four esophageal biopsies will be taken both from the distal (5 cm above the gastro-esophageal junction) and proximal (15 cm above the gastro-esophageal junction) esophagus. This number of biopsies has been shown to maximize the diagnostic sensitivity for EoE.
Histology and eosinophil counts:
All tissue samples from the Cytosponge and endoscopy will be coded with a subject's identification number, but will otherwise be masked for all clinical data, including EoE activity, symptoms, patient characteristics, and treatments prescribed. Using the paraffin blocks, pathology slides will be cut and the tissue processed with routine H&E staining. The slides will then be digitized, and using the Aperio ImageScope (Aperio Technologies, Vista, CA), the maximum eosinophil density (eosinophils/mm2 [eos/mm2]) will be determined using our previously validated protocol. For purposes of comparison to previous studies, eosinophil density will then be converted to eosinophil counts (eos/hpf) for an assumed hpf size of 0.24 mm2, the size of an average field as reported in the literature. The study pathologists from UNC and Mayo Clinic will review the specimens from their sites, and the study pathology from Cambridge will provide a second review of all specimens to ensure the most accurate quantification of eosinophil counts possible.
In addition, investigators plan to perform special staining and analysis of the existing biopsy and sponge samples with the goal of determining if the diagnostic accuracy of this test can be improved. In particular investigators will examine markers of eosinophil function, activation, and inflammation, such as eosinophil peroxidase (EPX), a granule protein that clearly identifies intact eosinophils, as well as extracellular EPX deposition suggestive of degranulation. This can be detected with immunohistochemistry. This would be done at Mayo clinic with our current collaborators who currently have the coded specimens.
Safety and accessibility assessments:
Patients will be assessed at multiple points to determine the safety of the Cytosponge in EoE. Investigators will assess for any symptoms or events as soon as the sponge capsule is swallowed, as well as immediately after the expanded sponge is removed. Participants will be contacted 1 and 7 days after the endoscopy to assess for adverse events. For Aim 2, participants will be administered the acceptability survey at the 7 day follow-up point, so patients have adequate time to reflect on their experiences with both tissue collection approaches. In particular this survey will record the patient's experience with swallowing the Cytosponge, whether they would do it again, and whether they prefer the Cytosponge or endoscopy for diagnosis and monitoring of EoE.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| EoE Patients | Experimental | Patients who have been diagnosed with EoE as per recent guidelines will be enrolled. Samples will be obtained using the Cytosponge™ Cell Collection Device (Cytosponge) prior to participants' routine endoscopy with biopsy. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Cytosponge™ Cell Collection Device | Device | The Cytosponge™ Cell Collection Device (Cytosponge) is intended to collect surface cells from the esophagus. The device consists of a swallowable capsule, which dissolves in the body cavity, releasing a self-expandable sponge. The sponge is then retrieved from the esophagus using an attached cord. During the retrieval process, the sponge collects cells from the most superficial layer of the esophageal mucosa. Once removed from the body cavity, the sponge and cells are retained for investigation and/or testing. The Cytosponge™ Cell Collection Device (Cytosponge) received 510(k) clearance from the FDA on November 26, 2014 (K142695). The Cytosponge ™ Cell Collection device is a Class II product under 21 CFR 874.4710 esophagoscope (flexible or rigid) and accessories. |
| Measure | Description | Time Frame |
|---|---|---|
| Percent Agreement Between Cytosponge and Endoscopic Biopsy Results | The primary outcome variables are sensitivity (percent agreement between positive results) and specificity (percent agreement between negative results) of the Cytosponge ability to detect the presence of EoE as compared to upper endoscopy with biopsy (the gold standard for diagnosis and monitoring of EoE). Overall agreement is defined as percentage of Cytosponge procedures yielding results consistent with endoscopic biopsy results. Presence of EoE is measured by the count of eosinophils present per high power field (eos/HPF) with active EoE defined as >=15 eos/HPF. Sensitivity was calculated via percentage of positive (active EoE) results obtained via Cytosponge as compared to results indicating active EoE via endoscopy with biopsy. Specificity was calculated via percentage of negative (inactive EoE) results obtained via Cytosponge as compared to results indicating inactive EoE via endoscopy with biopsy. | At study enrollment and initial procedure and each additional procedure, up to 1 year after enrollment |
| Measure | Description | Time Frame |
|---|---|---|
| Overall Agreement Between Cytosponge and Endoscopic Biopsy Results as Measured by Kappa | Overall agreement (Cytosponge procedures yielding results consistent with endoscopic biopsy results) as measured by Cohen's Kappa. Overall Cohen's Kappa is a statistical measure for assessing the reliability of agreement between the two results by taking into account the element of chance. Cohen's kappa can range from 0 to 1 with 1 indicating perfect agreement and 0 indicating an agreement equivalent to chance. |
| Measure | Description | Time Frame |
|---|---|---|
| Number of Responses Indicating Preference for Cytosponge Over Endoscopic Biopsy | The number of responses indicating preference for Cytosponge to endoscopic biopsy. Preference was measured by asking participants after each procedure, "which procedure would you prefer to undergo again if your physician indicated it was medically necessary?" with the options "Traditional Upper Endoscopy" and "Cytosponge." The total number of responses recorded as "Cytosponge" and the total number of responses recorded as "Traditional Upper Endoscopy" were summed. |
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Evan Dellon, MD, MPH | UNC-Chapel Hill | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Mayo Clinic | Rochester | Minnesota | 55902 | United States | ||
| University of North Carolina at Chapel Hill |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 22619364 | Background | Hirano I, Moy N, Heckman MG, Thomas CS, Gonsalves N, Achem SR. Endoscopic assessment of the oesophageal features of eosinophilic oesophagitis: validation of a novel classification and grading system. Gut. 2013 Apr;62(4):489-95. doi: 10.1136/gutjnl-2011-301817. Epub 2012 May 22. | |
| 16923475 | Background | Gonsalves N, Policarpio-Nicolas M, Zhang Q, Rao MS, Hirano I. Histopathologic variability and endoscopic correlates in adults with eosinophilic esophagitis. Gastrointest Endosc. 2006 Sep;64(3):313-9. doi: 10.1016/j.gie.2006.04.037. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | EoE Patients | Patients who have been diagnosed with EoE as per recent guidelines will be enrolled. Samples will be obtained using the Cytosponge™ Cell Collection Device (Cytosponge) prior to participants' routine endoscopy with biopsy. |
| Title | Milestones | Reasons Not Completed | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
|
Baseline characteristics unavailable for the 6 participants who were unable to swallow the Cytosponge.
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | EoE Patients | Patients who have been diagnosed with EoE as per recent guidelines will be enrolled. Samples will be obtained using the Cytosponge™ Cell Collection Device (Cytosponge) prior to participants' routine endoscopy with biopsy. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Continuous | Mean |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Percent Agreement Between Cytosponge and Endoscopic Biopsy Results | The primary outcome variables are sensitivity (percent agreement between positive results) and specificity (percent agreement between negative results) of the Cytosponge ability to detect the presence of EoE as compared to upper endoscopy with biopsy (the gold standard for diagnosis and monitoring of EoE). Overall agreement is defined as percentage of Cytosponge procedures yielding results consistent with endoscopic biopsy results. Presence of EoE is measured by the count of eosinophils present per high power field (eos/HPF) with active EoE defined as >=15 eos/HPF. Sensitivity was calculated via percentage of positive (active EoE) results obtained via Cytosponge as compared to results indicating active EoE via endoscopy with biopsy. Specificity was calculated via percentage of negative (inactive EoE) results obtained via Cytosponge as compared to results indicating inactive EoE via endoscopy with biopsy. | Of the 105 procedures performed on 80 participants, 101 yielded results from both Cytosponge and endoscopy with biopsy for comparison. | Posted | Number | Percent | At study enrollment and initial procedure and each additional procedure, up to 1 year after enrollment | Evaluable Specimens | Evaluable Specimens |
Adverse events were collected at each procedure (interventional procedure and specimen collection) through 30-days post-procedure, up to 1 year
Per study protocol section 7.2, only those events that are potentially related to participation in this research study were collected.
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | EoE Patients | Patients who have been diagnosed with EoE as per recent guidelines will be enrolled. Samples will be obtained using the Cytosponge™ Cell Collection Device (Cytosponge) prior to participants' routine endoscopy with biopsy. |
Not provided
Not provided
Due to device availability, this study was suspended indefinitely in June 2016. In August 2020, the decision was made to terminate the study early. The study population was limited to adult patients without critical strictures or small caliber esophagus.
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Evan Dellon, MD | University of North Carolina at Chapel Hill | 919-966-2511 | evan_dellon@med.unc.edu |
Not provided
| Type | Includes Protocol | Includes SAP | Includes ICF | Document Label | Document Date | Document Uploaded Date | Document File Name |
|---|---|---|---|---|---|---|---|
| Prot_SAP | Yes | Yes | No | Study Protocol and Statistical Analysis Plan | Sep 25, 2020 | Oct 21, 2020 | Prot_SAP_000.pdf |
Not provided
| ID | Term |
|---|---|
| D057765 | Eosinophilic Esophagitis |
| ID | Term |
|---|---|
| D004941 | Esophagitis |
| D004935 | Esophageal Diseases |
| D005767 | Gastrointestinal Diseases |
| D004066 | Digestive System Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
|
| At study enrollment and initial procedure and each additional procedure, up to 1 year after enrollment |
| Acceptability of Cytosponge Compared to Endoscopic Biopsy, as Measured by Visual Analog Scale | Acceptability of Cytosponge compared to endoscopic biopsy as measured by visual analogue scale. Participants were asked to rate their experience of the procedures on a scale of 0-10, where 0 indicates "unacceptable, very difficult even for a medical test," and 10 indicates "not an issue, would take test." A higher score indicates a more acceptable test. Acceptability was measured after each procedure and scores from each assessment were summed to obtain the mean and standard deviation. | 7 days after each procedure |
| Acceptability of Cytosponge as Measured by the Impact of Events Scale | Acceptability of Cytosponge as measured by the Impact of Events (IES) scale. The IES measures subjective distress (such as intrusive thoughts or emotions and avoidant or anxious behavior) following a stressful event. Respondents are asked to answer questions to indicate the amount of stress from the event. Scores are calculated using the following scale: Not at all =0, Rarely =1, Sometimes =3, Often =4. The total score is calculated by adding each response, with a total final score ranging from (0-60). Scores ranging 0-8 indicate no meaningful impact, scores ranging 9-25 indicate impact, and scores of 26 and above are considered very important (26-43 = powerful impact, 44-75 = severe impact). Acceptability was measured after each procedure and scores from each assessment were summed to obtain the mean and standard deviation. | 7 days after each procedure |
| 7 days after each procedure |
| Chapel Hill |
| North Carolina |
| 27516 |
| United States |
| 19830560 | Background | Dellon ES, Fritchie KJ, Rubinas TC, Woosley JT, Shaheen NJ. Inter- and intraobserver reliability and validation of a new method for determination of eosinophil counts in patients with esophageal eosinophilia. Dig Dis Sci. 2010 Jul;55(7):1940-9. doi: 10.1007/s10620-009-1005-z. Epub 2009 Oct 15. |
| 17617209 | Background | Dellon ES, Aderoju A, Woosley JT, Sandler RS, Shaheen NJ. Variability in diagnostic criteria for eosinophilic esophagitis: a systematic review. Am J Gastroenterol. 2007 Oct;102(10):2300-13. doi: 10.1111/j.1572-0241.2007.01396.x. Epub 2007 Jul 7. |
| 28809387 | Result | Katzka DA, Smyrk TC, Alexander JA, Geno DM, Beitia RA, Chang AO, Shaheen NJ, Fitzgerald RC, Dellon ES. Accuracy and Safety of the Cytosponge for Assessing Histologic Activity in Eosinophilic Esophagitis: A Two-Center Study. Am J Gastroenterol. 2017 Oct;112(10):1538-1544. doi: 10.1038/ajg.2017.244. Epub 2017 Aug 15. |
| years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Race (NIH/OMB) | Count of Participants | Participants |
|
| Region of Enrollment | Count of Participants | Participants |
|
| ID | Title | Description |
|---|---|---|
| OG000 | EoE Patients | Patients who have been diagnosed with EoE as per recent guidelines will be enrolled. Samples will be obtained using the Cytosponge™ Cell Collection Device (Cytosponge) prior to participants' routine endoscopy with biopsy. |
|
|
| Secondary | Overall Agreement Between Cytosponge and Endoscopic Biopsy Results as Measured by Kappa | Overall agreement (Cytosponge procedures yielding results consistent with endoscopic biopsy results) as measured by Cohen's Kappa. Overall Cohen's Kappa is a statistical measure for assessing the reliability of agreement between the two results by taking into account the element of chance. Cohen's kappa can range from 0 to 1 with 1 indicating perfect agreement and 0 indicating an agreement equivalent to chance. | Of the 105 procedures performed on 80 participants, 101 yielded evaluable specimens from both Cytosponge and endoscopy with biopsy for comparison. | Posted | Number | Kappa coefficient | At study enrollment and initial procedure and each additional procedure, up to 1 year after enrollment | Results of Evaluable Specimens | Results of Evaluable Specimens |
|
|
|
|
| Secondary | Acceptability of Cytosponge Compared to Endoscopic Biopsy, as Measured by Visual Analog Scale | Acceptability of Cytosponge compared to endoscopic biopsy as measured by visual analogue scale. Participants were asked to rate their experience of the procedures on a scale of 0-10, where 0 indicates "unacceptable, very difficult even for a medical test," and 10 indicates "not an issue, would take test." A higher score indicates a more acceptable test. Acceptability was measured after each procedure and scores from each assessment were summed to obtain the mean and standard deviation. | Posted | Mean | Standard Deviation | Units on a scale | 7 days after each procedure |
|
|
|
| Secondary | Acceptability of Cytosponge as Measured by the Impact of Events Scale | Acceptability of Cytosponge as measured by the Impact of Events (IES) scale. The IES measures subjective distress (such as intrusive thoughts or emotions and avoidant or anxious behavior) following a stressful event. Respondents are asked to answer questions to indicate the amount of stress from the event. Scores are calculated using the following scale: Not at all =0, Rarely =1, Sometimes =3, Often =4. The total score is calculated by adding each response, with a total final score ranging from (0-60). Scores ranging 0-8 indicate no meaningful impact, scores ranging 9-25 indicate impact, and scores of 26 and above are considered very important (26-43 = powerful impact, 44-75 = severe impact). Acceptability was measured after each procedure and scores from each assessment were summed to obtain the mean and standard deviation. | Not all participants competed the IES scale. Mean and Standard Deviation are reported for the 18 participants who completed the IES scale seven days after the procedure. | Posted | Mean | Standard Deviation | Units on a scale | 7 days after each procedure |
|
|
|
| Other Pre-specified | Number of Responses Indicating Preference for Cytosponge Over Endoscopic Biopsy | The number of responses indicating preference for Cytosponge to endoscopic biopsy. Preference was measured by asking participants after each procedure, "which procedure would you prefer to undergo again if your physician indicated it was medically necessary?" with the options "Traditional Upper Endoscopy" and "Cytosponge." The total number of responses recorded as "Cytosponge" and the total number of responses recorded as "Traditional Upper Endoscopy" were summed. | Data reported for participants who indicated a preference at follow up. | Posted | Number | responses | 7 days after each procedure |
|
|
|
| 0 |
| 86 |
| 0 |
| 86 |
| 0 |
| 86 |
Not provided
Not provided
| D005759 |
| Gastroenteritis |
| D004802 | Eosinophilia |
| D007960 | Leukocyte Disorders |
| D006402 | Hematologic Diseases |
| D006425 | Hemic and Lymphatic Diseases |
| D006969 | Hypersensitivity, Immediate |
| D006967 | Hypersensitivity |
| D007154 | Immune System Diseases |