The purpose of this study is to determine if subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) without training and obtain valid glucose results.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Intended Users of the Monitoring System | Experimental | Subjects with diabetes use the Karajishi Contour Investigational BG Monitoring System with no training. The criteria for the intended use population:
|
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Karajishi Contour Investigational BG Monitoring System | Device | Subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Karajishi Contour Investigational BG Monitoring System with no training. All BG results are compared to reference method results obtained from subject capillary plasma. Also, study staff test subject venous blood and BG results are compared to reference method results obtained from subject venous plasma. |
| Measure | Description | Time Frame |
|---|---|---|
| Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Subjects with diabetes self-test fingerstick blood use an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma). | 1 hour |
| Measure | Description | Time Frame |
|---|---|---|
| Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Study staff test subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results are used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI venous plasma) and +/-15% (>=100 mg/dL YSI venous plasma). |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Leslie J Klaff, MD | Rainier Clinical Research Center | Principal Investigator |
| Timothy Bailey, MD | AMCR Institute | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| AMCR Institute | Escondido | California | 92026 | United States | ||
| Rainier Clinical Research Center |
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Intended Users of the Monitoring System | Subjects with diabetes use the Karajishi Contour Investigational BG Monitoring System with no training. The criteria for the intended use population:
Karajishi Contour Investigational BG Monitoring System: Subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Karajishi Contour Investigational BG Monitoring System with no training. All BG results are compared to reference method results obtained from subject capillary plasma. Also, study staff test subject venous blood and BG results are compared to reference method results obtained from subject venous plasma. |
| Title | Milestones | Reasons Not Completed | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Intended Users of the Monitoring System | Subjects with diabetes use the Karajishi Contour Investigational BG Monitoring System with no training. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Customized | Mean |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Subjects with diabetes self-test fingerstick blood use an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma). | 135 (136-1) Blood glucose results were analyzed. One subject discontinued from testing after low BG fingerstick result (hypoglycemia AE), thus no reference result available. | Posted | Number | Blood glucose results | 1 hour |
|
1 hour subject visit
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Intended Users of the Monitoring System | Subjects with diabetes use the Karajishi Contour Investigational BG Monitoring System with no training. The criteria for the intended use population:
Karajishi Contour Investigational BG Monitoring System: Subjects with diabetes perform self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Karajishi Contour Investigational BG Monitoring System with no training. All BG results are compared to reference method results obtained from subject capillary plasma. Also, study staff test subject venous blood and BG results are compared to reference method results obtained from subject venous plasma. |
Not provided
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| Hypoglycemia | Endocrine disorders | Non-systematic Assessment | One subject discontinued from all testing after hypoglycemia AE. Subject's hypoglycemia was resolved at the visit. |
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Carmine Greene, Senior Clinical Research Scientist | Ascensia Diabetes Care | 574-257-3040 | carmine.greene@ascensia.com |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
| 1 hour |
| Number of Blood Glucose (BG) Results From Alternative Site Testing (AST) Palm Blood Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Subjects with diabetes self-test Alternative Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS AST palm results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma). | 1 hour |
| Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff | Study staff obtain and test subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results are used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma). | 1 hour |
| Number of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire Statements | Staff will obtain subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects may respond Strongly Agree; Agree; Neutral; Disagree; or Strongly Disagree. | 1 hour |
| Renton |
| Washington |
| 98057 |
| United States |
| years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Ethnicity (NIH/OMB) | Count of Participants | Participants |
|
| Race (NIH/OMB) | Subjects provided more than one choice for Race, therefore the apparent frequency total for Race was different from number of subjects. | Number | participants |
|
| Region of Enrollment | Number | participants |
|
|
|
| Secondary | Number of Venous Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Study staff test subject venous blood using an investigational Blood Glucose Monitoring System (BGMS). Venous BGMS results are compared with subject venous plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer venous plasma BG results are used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI venous plasma) and +/-15% (>=100 mg/dL YSI venous plasma). | 132 (136-4) Blood glucose results were analyzed. One subject discontinued from all testing after hypoglycemia AE. One (1) subject lab reference replicates were discrepant and not evaluable per protocol. Venipuncture was not successful for 2 subjects. | Posted | Number | Blood glucose results | 1 hour |
|
|
|
| Secondary | Number of Blood Glucose (BG) Results From Alternative Site Testing (AST) Palm Blood Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method | Subjects with diabetes self-test Alternative Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS) with no training. BGMS AST palm results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results are used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma). | 130 (136-6) Blood glucose results were analyzed. One subject discontinued from all testing after hypoglycemia AE. One (1) subject with low blood sugar did not attempt AST testing per protocol. No AST palm results were obtained for one (1) subject. Three (3) subjects had low blood sugar; AST results were not evaluable per protocol. | Posted | Number | Blood glucose results | 1 hour |
|
|
|
| Secondary | Number of Subject Fingerstick Blood Glucose (BG) Results Within +/- 15 mg/dL (<100 mg/dL) and Within +/- 15% (>=100 mg/dL) of Laboratory Glucose Method When Obtained and Tested by Study Staff | Study staff obtain and test subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results are compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results are used to calculate the number of BGMS results within +/-15 mg/dL (<100 mg/dL YSI capillary plasma) and +/-15% (>=100 mg/dL YSI capillary plasma). | 135 (136-1) Blood glucose results were analyzed. One subject discontinued from all testing after hypoglycemia AE. | Posted | Number | Blood glucose results | 1 hour |
|
|
|
| Secondary | Number of Subject Responses That Strongly Agree or Agree or Are Neutral With Questionnaire Statements | Staff will obtain subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects may respond Strongly Agree; Agree; Neutral; Disagree; or Strongly Disagree. | Posted | Number | participants | 1 hour |
|
|
|
| 0 |
| 136 |
| 4 |
| 136 |
|
Not provided
| Title | Measurements |
|---|---|
|
| I like the overall meter design. |
|
| I find the meter easy to use. |
|
| The instruction (User Guide) is easy to understand |
|
| Instructions clearly explain how to run a test. |
|
| Instructions clearly explain meter error messages. |
|