Energy expenditure (EE) in critically ill patients is highly variable depending on the initial severity of the disease and treatments. Clinicians need to measure EE by indirect calorimetry (IC) to optimize nutritional support. IC devices available on the market have insufficient accuracy for clinical and research use. A new IC is being developed to meet these needs.
Objectives: Validation of a reliable and easy-to-use IC device.
Primary objective:
• Accuracy of EE measurements (kcal/d measured over 30 min) of the new and the reference IC devices in intubated ICU patients on mechanical ventilation
Secondary objectives:
Design: Prospective, controlled, observational multicenter study.
Subjects: 182 ICU patients.
Energy expenditure (EE) in critically ill patients is highly variable depending on the initial severity of the disease and treatments. Clinicians need to measure EE by indirect calorimetry (IC) to optimize nutritional support. IC devices available on the market have insufficient accuracy for clinical and research use. A new IC is being developed to meet these needs.
Objectives: Validation of a reliable and easy-to-use IC device.
Primary objective:
• Accuracy of EE measurements (kcal/d measured over 30 min) of the new and the reference IC devices in intubated ICU patients on mechanical ventilation
Secondary objectives:
Inclusion criteria:
• Adult ICU patients with spontaneous breathing, or on invasive or non-invasive mechanical ventilation.
Exclusion criteria:
Outcomes:
Primary outcome:
• EE in intubated ICU patients on mechanical ventilation
Secondary outcomes:
Indirect calorimetry devices and methodology:
Barometric pressure is recorded. EE is calculated using the modified Weir equation.
Data collected:
Statistics and data analysis:
The Coordinating center analyses the data transferred by the Investigation centers
Statistical analysis significance is set at p<0.05.
For the subgroup of patients with prolonged EE measurements or on non-invasive ventilation:
Univariate and multivariate analysis are used to analyze the factors associated with observed variations.
Value of the study for clinical practice practice and research:
An accurate and reliable IC for nutrition specialists will allow to optimize energy prescription, a factor shown to improve the clinical outcome. An easy-to-use and affordable (10'000 $) IC will reduce both the burden on human resources and costs.
Study originality:
Agenda of actions:
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| indirect calorimetry measurement | Measurement of energy expenditure in made by indirect calorimetry in patients laying down in a supine position on a bed during ongoing measure by connecting the calorimeter to the endotracheal tube (invasive mechanical ventilation) or to the facemask (non-invasive mechanical ventilation), or by using a transparent canopy in plexiglas to cover the head while room-air flows through and is mixed with the expirate (spontaneous breathing). This mixture of patient breath and room air is finally collected and analyzed by the IC device for O2 and CO2 concentrations, and used to calculate EE. EE is calculated using the modified Weir equation. |
Not provided
| Measure | Description | Time Frame |
|---|---|---|
| calories reference IC devices in intubated ICU patients on mechanical ventilation | Accuracy of EE measurements (kcal/d measured over 30 min) of the new and the reference IC devices in intubated ICU patients on mechanical ventilation | 30 minutes |
| Measure | Description | Time Frame |
|---|---|---|
| calories | Accuracy of EE measurements in ICU patients of spontaneous breathing | 30 minutes |
| calories | Accuracy of EE measurement during ~30 min. vs. the sum of 6 periods of 60 min/12 hrs |
Not provided
Inclusion Criteria:
Exclusion Criteria:
FiO2 > 60%
Not provided
Not provided
Not provided
All adult ICU patients with spontaneous breathing, or on invasive or non-invasive mechanical ventilation admitted into the intesive care unit
| Name | Role | Phone | Extension | |
|---|---|---|---|---|
| Claude Pichard, Prof | Contact | +41 22 372 9349 | claude.pichard@unige.ch | |
| Claudia P Heidegger, MD | Contact | +41 22 372 7440 | claudia-paula.heidegger@hcuge.ch |
| Name | Affiliation | Role |
|---|---|---|
| Claude Pichard, Prof | University Hospital, Geneva | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Geneva University Hospital- Nutrition Unit | Recruiting | Geneva | 1211 | Switzerland |
Not provided
| OTHER |
| Karolinska University Hospital | OTHER |
Not provided
Not provided
Not provided
| 60 minutes |
| calories | Accuracy of EE measurements in ICU patients on non-invasive mechanical ventilation | 30 minutes |