Aim of the study: To validate accelerometers for continuous monitoring of global cardiac function during weaning from cardiopulmonary bypass (CPB) in cardiac surgery.
Hypothesis:
The following hypothesis will be tested: Accelerometer measures are clinical relevant indices of global left and right myocardial performance during weaning from cardiopulmonary bypass.
Model:
The performance of the accelerometer will be tested in a intraoperative setting, during weaning from cardiopulmonary bypass. Accelerometer measures of ventricular systolic and diastolic function will be compared to corresponding measures and indices of ventricular functions by echocardiography and hemodynamic monitoring during a step wise controlled weaning from cardiopulmonary bypass, ensuring precise control of pre- and afterload.
Not provided
Not provided
Not provided
Not provided
Not provided
| Measure | Description | Time Frame |
|---|---|---|
| Comparison of global heart function measures between accelerometer, echocardiography and hemodynamics. | Measures will be assessed at baseline before cardiopulmonary by pass, at different workloads during a stepwise weaning, and after the cannulas are removed. The accelerometer measurements will be compared to echocardiography and hemodynamics (stroke volume and cardiac output). Systolic global function will be assessed by comparing peak systolic velocity and displacement by the accelerometer and the corresponding findings in echocardiography. The accelerometer systolic measures will also be correlated to ejection fraction with echocardiography. Diastolic global function early inflow (E'), atrial inflow velocity (A') and E'/A' with accelerometers and echocardiography will be compared. | The patients will be followed during surgery, approximately 3 hours. |
Not provided
Not provided
The hypothesis will be tested in 20 elective patients.
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Patients with significant aortic stenosis scheduled for surgical AVR
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Erik Fosse, Professor | Oslo University Hospital, The Intervention Sentre | Study Director |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Oslo University Hospital, The Intervention Sentre | Oslo | 0027 | Norway |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 28364479 | Derived | Grymyr OH, Beitnes JO, Eidet J, Tollofsrud S, Fiane A, Skulstad H, Fosse E, Halvorsen PS. Detection of intraoperative myocardial dysfunction by accelerometer during aortic valve replacement. Interact Cardiovasc Thorac Surg. 2017 Feb 1;24(2):188-195. doi: 10.1093/icvts/ivw326. |