Clinical Investigational Protocol for 2 RBC Single Unit Recovery on MCS®+ 8150 Using LN832 Disposable Sets (US)
Clinical Investigational Protocol for 2 RBC Single Unit Recovery on MCS®+ 8150 Using LN832 Disposable Sets (US)
The purpose of this study is to validate that single units of RBCs collected by the Haemonetics MCS®+ 8150 system using LN832 disposable sets during an interrupted double red cell collection protocol meet all in vitro FDA requirements for non-leukoreduced red blood cells, and to provide data to support the 510(k) clearance of the MCS®+ 8150 system using LN832 disposable sets for this intended use.
This study will take place at two sites (Blood Center of Wisconsin (BCW) Milwaukee, WI and American Red Cross-Norfolk, VA).
A standard double red cell collection protocol will be initiated using the Haemonetics MCS®+ 8150 system using LN832 disposable sets. The collection procedure will be terminated after collection of one RBC unit. The maximum target red cell volume will be selected based on the donor predonation weight as specified in the nomogram included in the MCS®+ 8150 operation manual.
The RBC units will be stored for 42 days at 1° to 6°C.
Upon collection and at 42 days after collection, the RBC units will be weighed and tested for hematology parameters (white blood cell count, red blood cell count, residual white blood cell count, hemoglobin, hematocrit, platelet count, mean corpuscular volume, mean corpuscular hemoglobin, mean corpuscular hemoglobin concentration, red blood cell distribution width, and supernatant hemoglobin) and chemistry parameters (pH, potassium, glucose, lactate, and ATP).
Acceptance criteria for this study are as follows:
After collection of sixty (60) evaluable RBC units, the study will be considered supportive of the intended use noted above if zero (0) units fail the acceptance criteria.
If one (1) unit fails acceptance criteria after collection of sixty (60) evaluable RBC units, an additional seventy one (71) units will be collected.
The study will be considered supportive of the intended use noted above if not more than one (1) total unit out of 131 evaluable units fails the acceptance criteria.
If a total of two (2) units fail acceptance criteria, the study will not be considered supportive of the intended use noted above.
Inclusion Criteria:
Exclusion Criteria: