A Clinical Study to Evaluate the Performance of the C8 MediSensors Optical Glucose Monitor™ During On-Site Testing and Real Time Use Monitoring
A Clinical Study to Evaluate the Performance of the C8 MediSensors Optical Glucose Monitor™ During On-Site Testing and Real Time Use Monitoring
The purpose of this study is to compare glucose measurements from the C8 MediSensors non-invasive Optical Glucose Monitor™ System to blood glucose measurements from the Yellow Springs Instrument(YSI) blood glucose analyzer.
This is a prospective, non-randomized, open label study to evaluate the performance of the C8 MediSensors non-invasive Optical Glucose Monitor™ on individuals with diabetes mellitus. Capillary blood samples will be obtained throughout a daily study session; the study session may last for up to 12 hours. Blood glucose will be measured on a YSI blood glucose analyzer. Subjects may adjust food intake and insulin administration so device performance is evaluated across a broad range of glucose concentrations. Glucose measurements will be time matched and paired for comparison. Skin effects and adverse events will be evaluated.
Inclusion Criteria:
Exclusion Criteria: