This study will evaluate the accuracy of two rapid methods of measuring hemoglobin in patients with suspected hemorrhage. These methods will be compared with standard laboratory measurements.
Anemia and bleeding are major causes of morbidity and mortality in both surgical and nonsurgical patients. The current standard of care for monitoring patients at risk for bleeding is serial measurement of hemoglobin levels. At present, the photometric cyanmethemoglobin method is the most widely used technique for monitoring hemoglobin in the lab, and is currently the gold standard. However, this method has potential for delay before final results are obtained.
Immediate hemoglobin measurements are available with portable point-of-care devices such as the iSTAT, which can produce a measurement of hemoglobin concentration in less than 1 minute. Unfortunately, the accuracy of this device has been reported to vary with hemoglobin level, and as such may not be as accurate in detecting blood loss when compared with the gold standard of laboratory analysis.
Recently, a noninvasive, spectrophotometry-based monitoring technology has been developed. This novel technology measures the differential optical density of wavelengths of light passed through the finger in a method similar to conventional pulse oximetry. While some studies have reported that this device appears to be accurate in patients undergoing elective surgical procedures, more recent work suggests that this accuracy degrades with increased blood loss, lower oximeter signal quality and lower absolute Hgb values.
We will evaluate the accuracy of point-of-care and non-invasive SpHb measurements and utility of continuous hemoglobin monitoring in an intensive care unit setting. If these methods of rapid hemoglobin measurement can be validated in patients at risk for ongoing hemorrhage, use of this technology may result in earlier detection of ongoing hemorrhage, expedite appropriate treatment, and improve patient outcomes.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Patients with Suspected Hemorrhage | There is a single group of patients in this study -- those with suspected hemorrhage who satisfy the inclusion and exclusion criteria. The same set of measurements will be take from each patients and those measurements will be compared with one another to determine accuracy. |
Not provided
| Measure | Description | Time Frame |
|---|---|---|
| CBC Hemoglobin Measurement Compared to Non-invasive Radical-7 Measurement | Whenever blood was drawn for laboratory measurement of serum hemoglobin, we used one drop of blood to make point-of-care measurements using the CDC and Radical-7 methods. For purposes of reporting outcomes measures, we took an equally weighted average of measurements on each device (i.e., all measurement occasions on all patients). These are the means reported in the Outcome Measures Data Table. | n ≥ 1 measurements were taken each day. All measurements (n ≥ 7) from ICU Days 1-7 were used in Bland-Altman analysis, equally weighted. |
| CBC Hemoglobin Measurement Compared to Non-invasive iSTAT Measurement | When blood was drawn for laboratory measurement of serum hemoglobin, one drop of blood was used to make point of care measurements using the CBC and iSTAT methods. For purposes of reporting outcomes measures, we took an equally weighted average of measurements on each device (i.e., all measurement occasions on all patients). These are the means reported in the Outcome Measures Data Table. | n ≥ 1 measurements were taken each day. All measurements (n ≥ 7) from ICU Days 1-7 were used in Bland-Altman analysis, equally weighted. |
Not provided
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Patients admitted to the Surgical Intensive Care Unit of University Hospital, Cincinnati will be eligible for this study if they are at risk of ongoing bleeding and require serial CBC measurements. A patient will not be eligible if he or she is:(1) less than 18 years of age, (2) a prisoner, or if (3) clinicians are unable to take pulse oximetry readings due to injuries, burns, amputations, or related problems.
Not provided
Not provided
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| University Hospital | Cincinnati | Ohio | 45219 | United States |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 14453500 | Background | van KAMPEN E, ZIJLSTRA WG. Standardization of hemoglobinometry. II. The hemiglobincyanide method. Clin Chim Acta. 1961 Jul;6:538-44. doi: 10.1016/0009-8981(61)90145-0. No abstract available. | |
| 5608692 | Background | International committee for standardization in haematology. Recommendations for haemoglobinometry in human blood. Br J Haematol. 1967 Apr;13:71-5. doi: 10.1111/j.1365-2141.1967.tb00751.x. No abstract available. |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Patients With Suspected Hemorrhage | There is a single group of patients in this study -- those with suspected hemorrhage who satisfy the inclusion and exclusion criteria. The same set of measurements will be take from each patients and those measurements will be compared with one another to determine accuracy. |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Adult patients (>18 years of age) with the potential for hemorrhage and who were admitted to the Surgical Intensive Care Unit of a tertiary referral, Level I verified trauma center.
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Patients With Suspected Hemorrhage | There is a single group of patients in this study -- those with suspected hemorrhage who satisfy the inclusion and exclusion criteria. The same set of measurements will be take from each patients and those measurements will be compared with one another to determine accuracy. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | CBC Hemoglobin Measurement Compared to Non-invasive Radical-7 Measurement | Whenever blood was drawn for laboratory measurement of serum hemoglobin, we used one drop of blood to make point-of-care measurements using the CDC and Radical-7 methods. For purposes of reporting outcomes measures, we took an equally weighted average of measurements on each device (i.e., all measurement occasions on all patients). These are the means reported in the Outcome Measures Data Table. | Posted | Mean | 95% Confidence Interval | grams per deciliter | n ≥ 1 measurements were taken each day. All measurements (n ≥ 7) from ICU Days 1-7 were used in Bland-Altman analysis, equally weighted. |
|
During patient's hospital stay
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Radical-7 vs. CBC | All patients: Difference between Radical-Y hemoglobin measurement and CBC hemoglobin measurement |
Not provided
Not provided
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Betty Tsuei, MD | University of Cincinnati | 513-558-5661 | Betty.Tsuei@uc.edu |
Not provided
| ID | Term |
|---|---|
| D006470 | Hemorrhage |
| ID | Term |
|---|---|
| D010335 | Pathologic Processes |
| D013568 | Pathological Conditions, Signs and Symptoms |
Not provided
Not provided
Not provided
Not provided
Not provided
| 22238693 | Background | Gayat E, Aulagnier J, Matthieu E, Boisson M, Fischler M. Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies. PLoS One. 2012;7(1):e30065. doi: 10.1371/journal.pone.0030065. Epub 2012 Jan 6. |
| 22116696 | Background | May JM. Vitamin C transport and its role in the central nervous system. Subcell Biochem. 2012;56:85-103. doi: 10.1007/978-94-007-2199-9_6. |
| 21640577 | Background | Butwick AJ, Hilton G, Riley ET, Carvalho B. Non-invasive measurement of hemoglobin during cesarean hysterectomy: a case series. Int J Obstet Anesth. 2011 Jul;20(3):240-5. doi: 10.1016/j.ijoa.2011.03.009. Epub 2011 Jun 2. |
| 21039345 | Background | Hahn RG, Li Y, Zdolsek J. Non-invasive monitoring of blood haemoglobin for analysis of fluid volume kinetics. Acta Anaesthesiol Scand. 2010 Nov;54(10):1233-40. doi: 10.1111/j.1399-6576.2010.02321.x. |
| 19951386 | Background | Myers D, McGraw M, George M, Mulier K, Beilman G. Tissue hemoglobin index: a non-invasive optical measure of total tissue hemoglobin. Crit Care. 2009;13 Suppl 5(Suppl 5):S2. doi: 10.1186/cc8000. Epub 2009 Nov 30. |
| 19951382 | Background | Santora RJ, Moore FA. Monitoring trauma and intensive care unit resuscitation with tissue hemoglobin oxygen saturation. Crit Care. 2009;13 Suppl 5(Suppl 5):S10. doi: 10.1186/cc8008. Epub 2009 Nov 30. |
| 21716091 | Background | Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011 Sep;115(3):548-54. doi: 10.1097/ALN.0b013e3182270c22. |
| 22133758 | Background | Applegate RL 2nd, Barr SJ, Collier CE, Rook JL, Mangus DB, Allard MW. Evaluation of pulse cooximetry in patients undergoing abdominal or pelvic surgery. Anesthesiology. 2012 Jan;116(1):65-72. doi: 10.1097/ALN.0b013e31823d774f. |
| 25159346 | Derived | Tsuei BJ, Hanseman DJ, Blakeman MJ, Blakeman TC, Yang SH, Branson RD, Gerlach TW. Accuracy of noninvasive hemoglobin monitoring in patients at risk for hemorrhage. J Trauma Acute Care Surg. 2014 Sep;77(3 Suppl 2):S134-9. doi: 10.1097/TA.0000000000000326. |
| Participants |
|
| Sex: Female, Male | Count of Participants | Participants |
|
CBC hemoglobin measurement |
|
|
|
| Primary | CBC Hemoglobin Measurement Compared to Non-invasive iSTAT Measurement | When blood was drawn for laboratory measurement of serum hemoglobin, one drop of blood was used to make point of care measurements using the CBC and iSTAT methods. For purposes of reporting outcomes measures, we took an equally weighted average of measurements on each device (i.e., all measurement occasions on all patients). These are the means reported in the Outcome Measures Data Table. | Posted | Mean | 95% Confidence Interval | grams per deciliter | n ≥ 1 measurements were taken each day. All measurements (n ≥ 7) from ICU Days 1-7 were used in Bland-Altman analysis, equally weighted. |
|
|
|
|
| 0 |
| 88 |