Verify the Functions and Efficiency of Transtek Glass Body Fat Analyzer, GBF-733-W1, GBF-1012, BF-1039, BF-1041-A, GBF-950-D.
Verify the Functions and Efficiency of Transtek Glass Body Fat Analyzer, GBF-733-W1, GBF-1012, BF-1039, BF-1041-A, GBF-950-D.
The clinical protocol of the clinical testing of this device:
Hospital Information The data was collected by Zhongshan City People's Hospital Clinical Investigator Team at Zhongshan City People's Hospital, No.2 Sunwen Dong Road, Zhongshan, Guangdong 528403, P. R. China.
Investigator: Dr. Stephen Feng, Leader; Fan Futao, Nurse A; Wu Yanting, Nurse B.
Contact Dr. James Zhu Tel: +86 760 88823818 Used Equipments DUT (Device Under Test): Transtek Glass Body Fat Analyzer, Model: GBF-733-W1, GBF-1012, BF-1039, BF-1041-A, and GBF-950-D.
Reference Device: Transtek Glass Body Analyzer, GBF-950 (Predicate Device). Test Protocol
Test Purpose:
The aim of clinical test is to collect weight, body fat, total body water, bone mass, and muscle mass reading value which measured by DUT and reference device.
Note: GBF-733-W1 and BF-1039 have not bone mass functions. So it just collects weight, body fat, total body water, and muscle mass data.
Target Subject:
The subject database shall contain at least 25 male and 25 female subjects and all above 18 years old. Patient's population analysis is bellow.
Test procedures 1) Record the name, age, and gender values of each subject. 2) Let subject seated calmly for two minutes before test, dry the feet. 3) Use device to measure weight, body fat, total body water, bone mass, and muscle mass; record reading value.
4) For each subject, repeat 3) to get 3 pairs measurements. 5) Repeat step 1) to 4) for every device (DUT: GBF-733-W1, GBF-1012, BF-1039, BF-1041-A, and GBF-950-D; and reference device: GBF-950).
6) Repeat step 1) to 5) for each subject. 5. Note:
Inclusion Criteria:
Exclusion Criteria: