Method Comparison - Sample Inclusion Criteria
The following are criteria for inclusion of extracted genomic DNA samples in the method comparison sample pool:
- Sample is from a patient referred for post-natal cytogenetic testing.
- Sample gender is known.
- Sample quantity available for testing is ≥ 1 microgram of genomic DNA at a concentration of 60-80 nanograms per microliter (60-80ng/µl).
- Sample is extracted genomic DNA from EDTA or heparin-anticoagulated whole blood.
- Sample has been stored at 2 to 8°C or -15 to -25°C for no greater than three years from the date of extraction.
- Sample has been tested by a reference method. Acceptable reference methods include karyotype, FISH, qPCR, MLPA, and methylation analysis.
Method Comparison - Sample Exclusion Criteria
- Sample is from a patient not referred for post-natal cytogenetic testing.
- Sample is from a patient referred for cytogenetic oncology testing.
- Sample quantity < 1 microgram of genomic DNA or less than 60 nanograms per microliter (60ng/µl).
- Sample was improperly stored or was extracted from a sample that was improperly stored.
- Sample was tested by an Illumina array during standard of care testing.
Clinical Specificity - Sample Inclusion Criteria
The following are criteria for inclusion of extracted genomic DNA samples in the clinical specificity sample pool:
- Sample gender is known.
- Sample is from a patient not referred for post-natal cytogenetic testing.
- Sample quantity available for testing is ≥ 1 microgram of genomic DNA at a concentration of at least 60-80 nanograms per microliter (60-80ng/µl).
- Sample is extracted genomic DNA from EDTA or heparin-anticoagulated whole blood.
- Sample has been stored at 2 to 8°C or -15 to -25°C for no greater than three years from the date of extraction.
Clinical Specificity - Sample Exclusion Criteria
- Sample is from a patient referred for post-natal cytogenetic testing.
- Sample quantity < 1 microgram of genomic DNA or less than 60 nanograms per microliter (60ng/µl).
- Sample was improperly stored or was extracted from a sample that was improperly stored.