The purpose of this study is to evaluate the performance of an investigational blood glucose meter with the Contour sensor strip. It was decided that this investigational monitoring system would not be part of a regulatory submission and it will not proceed to marketed product.
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Intended Users of the System | Experimental | Untrained subjects with diabetes obtained capillary fingerstick, palm, and forearm blood and performed Blood Glucose (BG) tests using a Ninja 2 investigational blood glucose meter and the ContourĀ® sensor. This BG monitoring system will not proceed to marketed product. |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Ninja 2 Investigational Blood Glucose Meter | Device | Untrained subjects with diabetes performed Blood Glucose(BG) tests from the subject's capillary blood obtained from fingerstick, palm, and forearm using the Ninja 2 investigational meter. All BG results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff. |
| Measure | Description | Time Frame |
|---|---|---|
| Numbers of Fingerstick Blood Glucose (BG) Results Within +/- 5 to 15mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method | Subjects with diabetes tested subject fingerstick blood using an investigational blood glucose meter (BGM). BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5 to 20% (for reference BG results >=75mg/dL) of the reference method results. | 2 hours |
| Measure | Description | Time Frame |
|---|---|---|
| Numbers of Palm Blood Glucose (BG) Results Within +/- 5 to 15 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method | Subjects with diabetes used an investigational blood glucose meter (BGM) with subject capillary blood obtained from the palm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results. |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Michael Caswell, PhD | Consumer Product Testing Company | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Consumer Product Testing Company | Fairfield | New Jersey | 07004 | United States |
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Intended Users of the System | Untrained subjects with diabetes obtained capillary fingerstick, palm, and forearm blood and performed Blood Glucose (BG) tests using a Ninja 2 investigational blood glucose meter and the ContourĀ® sensor. Ninja 2 Investigational Blood Glucose Meter : Untrained subjects with diabetes performed Blood Glucose(BG) tests from the subject's capillary blood obtained from fingerstick, palm, and forearm using the Ninja 2 investigational meter. All BG results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff. |
| Title | Milestones | Reasons Not Completed | |||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Intended Users of the System | Untrained subjects with diabetes obtained capillary fingerstick, palm, and forearm blood and performed Blood Glucose (BG) tests using a Ninja 2 investigational blood glucose meter and the ContourĀ® sensor. Ninja 2 Investigational Blood Glucose Meter : Untrained subjects with diabetes performed Blood Glucose(BG) tests from the subject's capillary blood obtained from fingerstick, palm, and forearm using the Ninja 2 investigational meter. All BG results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff. |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Continuous | Mean |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Numbers of Fingerstick Blood Glucose (BG) Results Within +/- 5 to 15mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method | Subjects with diabetes tested subject fingerstick blood using an investigational blood glucose meter (BGM). BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5 to 20% (for reference BG results >=75mg/dL) of the reference method results. | 178 BG results possible-88 subjects tested 2 strip lots(88x2)/2 subjects had 1 reading(2x1). Withdrawn by Principal Investigator(PI)-3 subjects experienced Adverse Events before performance testing / one subject did not meet inclusion/exclusion. Data from 4 subjects were not evaluable since exceeded time between meter testing and reference method. | Posted | Number | Number of BG Test Results | 2 hours | BG test results | Participants |
|
Not provided
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Intended Users of the System | Untrained subjects with diabetes obtained capillary fingerstick, palm, and forearm blood and performed Blood Glucose (BG) tests using a Ninja 2 investigational blood glucose meter and the ContourĀ® sensor. Ninja 2 Investigational Blood Glucose Meter : Untrained subjects with diabetes performed Blood Glucose(BG) tests from the subject's capillary blood obtained from fingerstick, palm, and forearm using the Ninja 2 investigational meter. All BG results were compared to a reference laboratory glucose method. Subjects' success at performing basic tasks using only the User Guide were rated by study staff. |
Not provided
Not provided
It was decided that this investigational BG monitoring system would not be part of a regulatory submission. It will not proceed to market.
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Carmine Greene, Senior Clinical Research Scientist | Ascensia Diabetes Care | 914-333-6280 | carmine.greene@ascensia.com |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
|
| 2 hours |
| Number of Subjects Rated as<=3 Performing Basic Meter Tasks (Labeling Comprehension) | Subjects reviewed the instructions for use (User Guide and Quick Reference Guide) to learn to use the system. Study staff then observed and rated the subjects(1 to 4) on their success at performing the tasks. Scale: 1.Success in performing tasks correctly without assistance. 2.Successful with additional review of User Guide. 3.Successful with additional review and study staff assist similar to review of a specific function during a Customer Service call. 4.Subject did not perform task correctly and study staff intervention was required. | 2 hours |
| Numbers of Forearm Blood Glucose (BG) Results Within +/- 5 to 15 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method | Subjects with diabetes used an investigational blood glucose meter (BGM) with subject capillary blood obtained from the forearm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results. | 2 hours |
| years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Region of Enrollment | Number | participants |
|
| Intended Users of the System |
Untrained subjects with diabetes obtained capillary fingerstick, palm, and forearm blood and performed Blood Glucose (BG) tests using a Ninja 2 investigational blood glucose meter and the ContourĀ® sensor. |
|
|
| Secondary | Numbers of Palm Blood Glucose (BG) Results Within +/- 5 to 15 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method | Subjects with diabetes used an investigational blood glucose meter (BGM) with subject capillary blood obtained from the palm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results. | 161 BG results-80 subjects tested 2 strip lots;1 subject had 1 reading. Withdrawn by PI- 3 subject Adverse Events before performance testing;1 subject did not meet inclusion/exclusion. 6 subjects- Data not evaluable:exceeded time between meter testing and reference method. 7 subjects-Low bg results disallowed palm testing per protocol. | Posted | Number | Number of BG Test Results | 2 hours | Number of BG test results | Participants |
|
|
|
| Secondary | Number of Subjects Rated as<=3 Performing Basic Meter Tasks (Labeling Comprehension) | Subjects reviewed the instructions for use (User Guide and Quick Reference Guide) to learn to use the system. Study staff then observed and rated the subjects(1 to 4) on their success at performing the tasks. Scale: 1.Success in performing tasks correctly without assistance. 2.Successful with additional review of User Guide. 3.Successful with additional review and study staff assist similar to review of a specific function during a Customer Service call. 4.Subject did not perform task correctly and study staff intervention was required. | 97 participants analyzed: one subject was found not to meet inclusion/exclusion criteria so subject was withdrawn and no data from this subject was evaluated. For palm and forearm testing tasks, 2 additional subjects (with nonevaluable BG data) were not included. Only 95 participants were rated for those particular tasks. | Posted | Number | participants | 2 hours |
|
|
|
| Secondary | Numbers of Forearm Blood Glucose (BG) Results Within +/- 5 to 15 mg/dL (<75 mg/dL) or Within +/- 5% to 20% (>=75 mg/dL) of Laboratory Glucose Method | Subjects with diabetes used an investigational blood glucose meter (BGM) with subject capillary blood obtained from the forearm. BGM results were compared to a reference lab glucose method - Yellow Springs Instrument (YSI) Analyzer. BG results were used to calculate the number of BG results within +/- 5 to 15 mg/dL (for reference BG results <75mg/dL) or +/- 5% to 20% (for reference BG results >=75mg/dL) of the reference method results. | 158 BG results-78 subjects tested 2 strip lots;2 subjects tested 1 strip lot. 7 subjects-Low BG results disallowed forearm testing per protocol. 1 subject-Missing data. Remainder same as 'palm' analysis population description. | Posted | Number | Number of BG Test Results | 2 hours | Number of BG Test Results | Participants |
|
|
|
| 0 |
| 98 |
| 0 |
| 98 |
Not provided
|
| # within +/-5mg/dL (<75mg/dL) or 5% (>=75mg/dL) |
|
| Title | Measurements |
|---|---|
|
| Subject blood testing - fingerstick |
|
| Subject blood testing - palm (of 95) |
|
| Subject blood testing - forearm (of 95) |
|
|