The purpose of this study is to assess the performance of the subcutaneous glucose sensor over an extended sensor life. The sensor is currently approved for 3 days of use and this study will examine sensor safety and accuracy when used for six days. This study will also test sensor safety and accuracy when inserted in an alternate body location (buttock area in addition to abdomen area). The study hypothesis is that the sensor performance will not greatly diminish when used for six days, or in an alternate insertion area.
The Medtronic MiniMed Subcutaneous Glucose Sensor was originally approved by the FDA for commercialization as part of the Continuous Glucose Monitoring System (CGMS) on June 15, 1999 (PMA 980022). The Sensor is composed of a microelectrode with a thin coating of glucose oxidase beneath several layers of biocompatible membrane. This same sensor is used as part of the Guardian REAL-Time System, the latest advance in continuous glucose monitoring, which is based on the CGMS. Similar to the CGMS, the Guardian REAL-Time System has been developed for use in conjunction with a standard home blood glucose meter. The Guardian REAL-Time received regulatory approval from the FDA in 2006.
As currently used, the Subcutaneous Glucose Sensor is labeled for a maximum use duration of 72 hours, using only the abdomen area as an insertion site. Recent studies have shown that the useful sensor life could extend beyond three days, and it is reasonable to expect a significant percentage of sensors to last six days. It is the goal of this study to confirm sensor performance accuracy data from one of these recent studies. The sensor is also commonly worn in body areas other than the abdomen (such as the buttock). This study will also demonstrate sensor accuracy when used in an alternate site.
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| MMT-7003 subcutaneous glucose sensor | Device | All subjects will use the glucose sensor, randomized to participate in a 12-hour frequent (15-minute) blood glucose sampling period on one of six days of sensor wear. Blood samples will be analyzed using a laboratory standard (YSI) for comparison to the glucose sensor to evaluate accuracy. |
|
| Measure | Description | Time Frame |
|---|---|---|
| Glucose Sensor Accuracy When Compared to Laboratory Standard (YSI-Yellow Springs Instruments) | The primary accuracy parameter (primary effectiveness endpoint) was the comparative readings of paired sensor and YSI glucose readings, measured on days 1 through 6. Accuracy is defined as within 20% agreement between YSI and paired sensor (within 20 mg/dL if YSI <80 mg/dL). Accuracy ranges from 0 - 100, with higher number suggests better accuracy. | Days one through six of sensor use |
| Measure | Description | Time Frame |
|---|---|---|
| Device Related Moderate or Device Related Severe Adverse Events | Device related moderate adverse event: low level of inconvenience or concern to the subject and may interfere with daily activities but is usually improved by simple therapeutic remedy Device related severe adverse event: interrupts a subject's daily activity and typically requires intervening treatment. Note: device related determination is made by the site that there is a reasonable possibility that the adverse event may have been caused by the device. |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| John Mastrototaro, PHD | Medtronic MiniMed, Inc. | Study Director |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| AMCR Institute | Escondido | California | 92026 | United States | ||
| Sansum Diabetes Research Institute |
Not provided
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Subcutaneous Sensor Group | All subjects wore subcutaneous sensors |
| Title | Milestones | Reasons Not Completed | ||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Subcutaneous Sensor Group | All subjects wore subcutaneous sensors |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | |||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Glucose Sensor Accuracy When Compared to Laboratory Standard (YSI-Yellow Springs Instruments) | The primary accuracy parameter (primary effectiveness endpoint) was the comparative readings of paired sensor and YSI glucose readings, measured on days 1 through 6. Accuracy is defined as within 20% agreement between YSI and paired sensor (within 20 mg/dL if YSI <80 mg/dL). Accuracy ranges from 0 - 100, with higher number suggests better accuracy. | 61 subjects of 63 enrolled subjects (a total of 4971 paired YSI and sensor readings) completed participation in the inpatient frequent blood sampling procedure. | Posted | Number | 95% Confidence Interval | paired YSI/sensor glucose values | Days one through six of sensor use | paired YSI/sensor glucose values | Participants |
|
Not provided
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Subcutaneous Sensor Group | All subjects wore subcutaneous sensors |
Not provided
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| Broken hand | Injury, poisoning and procedural complications | Systematic Assessment | The 1 adverse event reported was a broken hand (right hand, fourth and fifth metacarpal fractured) which occurred when the subject intentionally "punched a truck." |
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Scott Lee M.D. | Medtronic | 818-576-4204 | scott.w.lee@medtronic.com |
Not provided
| ID | Term |
|---|---|
| D003920 | Diabetes Mellitus |
| ID | Term |
|---|---|
| D044882 | Glucose Metabolism Disorders |
| D008659 | Metabolic Diseases |
| D009750 | Nutritional and Metabolic Diseases |
| D004700 | Endocrine System Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| days one through six of sensor wear |
| Santa Barbara |
| California |
| 93105 |
| United States |
| Rainier Clinical Research Center | Renton | Washington | 98057 | United States |
| Participants |
|
| Age Continuous | Mean | Standard Deviation | years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Region of Enrollment | Number | participants |
|
|
|
|
| Secondary | Device Related Moderate or Device Related Severe Adverse Events | Device related moderate adverse event: low level of inconvenience or concern to the subject and may interfere with daily activities but is usually improved by simple therapeutic remedy Device related severe adverse event: interrupts a subject's daily activity and typically requires intervening treatment. Note: device related determination is made by the site that there is a reasonable possibility that the adverse event may have been caused by the device. | Posted | Number | events | days one through six of sensor wear |
|
|
|
| 0 |
| 63 |
| 1 |
| 63 |
|
Not provided