ISSUE 3 is a multi-center, prospective, randomised controlled double-blind study aimed to assess the effectiveness of pacemaker therapy for prevention of asystolic neurally-mediated syncope.
In asystolic neurally-mediated syncope (NMS) documented by Implantable Loop recorder (ILR), ISSUE-2, an observational trial, showed that pacemaker was effective in reducing the 1-year first syncope recurrence rate from 33% rate before implant (ILR phase 1) to 5% rate after implant (phase 2). Moreover, the control non-asystolic group still continued to have a 41% recurrence rate after the first recurrence of syncope, thus supporting the conclusion that the reduction with pacemaker was due to the beneficial effect of pacemaker itself and not to other factors. However a formal controlled trial is needed to confirm these findings.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Dual chamber pacemaker | Placebo Comparator | Dual chamber pacemaker programmed ODO (switched OFF) |
|
| Dual chamber pacemeker | Active Comparator | Medtronic dual chamber pacemaker programmed ON and with Rate Drope Response programmed ON |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Dual chamber pacemeker | Device |
|
|
| Measure | Description | Time Frame |
|---|---|---|
| Syncope Recurrence Rate | Intention to treat analysis of percentage of patients with syncope recurrence at 2 years follow-up after study arm assignement | 2 years |
Not provided
Not provided
Inclusion Criteria:
Exclusion criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Michele Brignole, MD | Ospedali del Tigullio, Lavagna | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Medtronic Italia S.p.A. | Rome | 00165 | Italy |
| PubMed Identifier | Type | Citation | Retractions |
|---|---|---|---|
| 22565936 | Derived | Brignole M, Menozzi C, Moya A, Andresen D, Blanc JJ, Krahn AD, Wieling W, Beiras X, Deharo JC, Russo V, Tomaino M, Sutton R; International Study on Syncope of Uncertain Etiology 3 (ISSUE-3) Investigators. Pacemaker therapy in patients with neurally mediated syncope and documented asystole: Third International Study on Syncope of Uncertain Etiology (ISSUE-3): a randomized trial. Circulation. 2012 May 29;125(21):2566-71. doi: 10.1161/CIRCULATIONAHA.111.082313. Epub 2012 May 7. |
| Label | URL |
|---|---|
| ISSUE3 main results publication PubMed | View source |
Not provided
12 out of 89 eligible pts refused randomization
From April 2007 to April 2011, 89 out of 511 patients (pts) implanted with implantable loop recorder (ILR) documented an asystolic syncope of >=3 sec or a non-syncopal asystole of >= 6 sec, inclusion criteria to be implanted with a dual chamber pacemaker (PM) and randomized to PM on or PM off. 77 pts out of 89 eligible pts were randomized.
Not provided
| ID | Title | Description |
|---|---|---|
| FG000 | Dual Chamber Pacemaker OFF | Implant dual chamber pacemaker programmed ODO (switched OFF) |
| FG001 | Dual Chamber Pacemeker ON | Implant dual chamber pacemaker programmed ON and with Rate Drope Response algorhythm programmed ON |
| Title | Milestones | Reasons Not Completed | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Overall Study |
|
Not provided
Not provided
| ID | Title | Description |
|---|---|---|
| BG000 | Dual Chamber Pacemaker OFF | Implant dual chamber pacemaker programmed ODO (switched OFF) |
| BG001 | Dual Chamber Pacemeker ON | Implant dual chamber pacemaker programmed ON and with Rate Drope Response algorhythm programmed ON |
| Units | Counts |
|---|---|
| Participants |
|
| Title | Description | Population Description | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Denominator Units Selected | Denominators | Classes |
|---|---|---|---|---|---|---|---|---|---|
| Age, Categorical | Count of Participants |
| Type | Title | Description | Population Description | Reporting Status | Anticipated Posting Date | Parameter Type | Dispersion Type | Unit of Measure | Calculate Percentage | Time Frame | Units Analyzed | Denominator Units Selected | Arm/Group Information | Denominators | Classes | Analyses | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Primary | Syncope Recurrence Rate | Intention to treat analysis of percentage of patients with syncope recurrence at 2 years follow-up after study arm assignement | 77 patients implanted with dual chamber pacemaker: 39 patients randomized to pacemaker OFF and 38 patients randomized to pacemaker ON | Posted | Number | 95% Confidence Interval | percentage of participants | 2 years |
|
Not provided
Not provided
Not provided
| ID | Title | Description | Deaths (Affected) | Deaths (At Risk) | Serious Events (Affected) | Serious Events (At Risk) | Other Events (Affected) | Other Events (At Risk) |
|---|---|---|---|---|---|---|---|---|
| EG000 | Dual Chamber Pacemaker OFF | Implant dual chamber pacemaker programmed ODO (switched OFF) |
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| Right ventricle lead dislodgment | Surgical and medical procedures | Systematic Assessment |
| Term | Organ System | Source Vocabulary | Assessment Type | Notes | Statistical Information |
|---|---|---|---|---|---|
| Atrial arrhythmia | Cardiac disorders | Systematic Assessment |
Not provided
| Title | Organization | Phone | Extension | |
|---|---|---|---|---|
| Silvia Giuli | Medtronic Italia SpA | 003932814224 | silvia.giuli@medtronic.com |
Not provided
| ID | Term |
|---|---|
| D013575 | Syncope |
| ID | Term |
|---|---|
| D014474 | Unconsciousness |
| D003244 | Consciousness Disorders |
| D019954 | Neurobehavioral Manifestations |
| D009461 | Neurologic Manifestations |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| BG002 | Total | Total of all reporting groups |
| Participants |
|
| Age, Continuous | Mean | Standard Deviation | years |
|
| Sex: Female, Male | Count of Participants | Participants |
|
| Region of Enrollment | Number | participants |
|
|
|
|
| 3 |
| 39 |
| 0 |
| 39 |
| EG001 | Dual Chamber Pacemeker ON | Implant dual chamber pacemaker programmed ON and with Rate Drope Response algorhythm programmed ON | 3 | 38 | 2 | 38 |
| Right atrium lead dislodgment | Surgical and medical procedures | Systematic Assessment |
|
| Subclavian vein thrombosis | Surgical and medical procedures | Systematic Assessment |
|
| Dead | Neoplasms benign, malignant and unspecified (incl cysts and polyps) | Systematic Assessment |
|
| Gastroenteritis | Gastrointestinal disorders | Systematic Assessment |
|
Not provided
Not provided
| D009422 | Nervous System Diseases |
| D012816 | Signs and Symptoms |
| D013568 | Pathological Conditions, Signs and Symptoms |