Not provided
The purpose of this study is to compare the early migration of two uncemented total knee arthroplasties with different metal coating. Only the tibial prosthesis plateau varies. One is made out of tantalum and has two short pegs for fixation into the tibia while the other is covered by titanium fiber-mesh on the bone-near site and has 4 short pegs for screw fixation into the tibia.
Around 4% of total knee arthroplasties (TKA) are revised 15 years after the primary operation, most due to loosening of the implant. The use of uncemented prostheses are expected to extend the fixation time and thereby the years of TKA survival. When a cementless implant is inserted, direct bone ingrowth to the implant depends on a complete primary stabilisation. Movement between the bone and implant will cause a fibrous membrane leading to micromotions. Micromotions can be evaluated by radiophotogrammetric analysis (RSA) in early follow-up and they have been shown to be a prognostic marker for painful loosening of the implant.
Tantalum is a fairly new biomaterial designed with a high and bone-binding porosity. Tantalum has already proven to be a good prosthesis material facilitating fast bone-ingrowth and supplying mechanical qualities in the bone-prosthesis interface that surpasses the qualities of regular prosthesis materials by uncemented method. Titanium on the other hand has been used as a prosthesis coating for years and also promote prosthesis fixation by osteointegration.
The purpose of this study is to compare the early migration of two uncemented total knee implants with different metal coating and fixation method. Only the tibial prosthesis plateau varies in design. One plateau is a 2 pegged tantalum monobloc and the other plateau is a modular titanium fiber-mesh coating with 4 short pegs for screw fixation into the tibia.
Furthermore the extend of the expected micromovements between the polyethylene liner and the metal backing of the titanium plateau will be compared to the stable tantalum monobloc.
Theoretically tantalum should allow a more optimal coating for bone ingrowth, but as primary stability is also important for longterm prostheses survival the method of primary fixation could alså play a role.
To evaluate both the plateau-bone micromotions and the polyethylene-prosthesis micromovements metal markers have been placed on the prosthesis pre-operatively and furthermore into the tibial bone and into the tibial polyethylene during surgery allowing us to evaluate stereo x-rays by a photogrammetric analyses called RSA at the Orthopaedic Center, Aarhus University Hospital. Follow-up stereo x-rays will be scheduled for 1 week, 3 months, 6 month, 1 years, 2 years and 5 years post-surgery.
Not provided
Not provided
Not provided
Not provided
| Label | Type | Description | Intervention Names |
|---|---|---|---|
| Tantalum knee | Active Comparator | Tantalum Tibial component, uncemented |
|
| Titanium Knee | Active Comparator | Titanium Tibial Component, screw fixed |
|
| Name | Type | Description | Arm Group Labels | Other Names |
|---|---|---|---|---|
| Tibial component | Device | Tibial component inserted at the time of surgery |
|
| Measure | Description | Time Frame |
|---|---|---|
| Polyethylene wear (mm) of the medial and lateral knee prosthesis compartments at 5 years followup measured with model-based radiostereometry | Up to 5 years |
| Measure | Description | Time Frame |
|---|---|---|
| Difference in MPTM (mm) tibial component migration from 2-5 years (stabilization 3) measured with model-based radiostereometry | up to 5 years | |
| Number of implants migrating above 0.2mm MTPM from 1-2 years (stabilization 2) and 2-5 years (stabilization 3) measured with model-based radiostereometry |
Not provided
Inclusion Criteria:
Exclusion Criteria:
Not provided
Not provided
Not provided
Not provided
Not provided
| Name | Affiliation | Role |
|---|---|---|
| Kjeld Søballe, MD, Prof. | Orthopaedic Center, Aarhus University Hospital, Palle Juul-Jensens Boulevard 165, 8200 Aarhus N, Denmark | Principal Investigator |
| Facility | Status | City | State | ZIP | Country | Contacts |
|---|---|---|---|---|---|---|
| Orthopaedic Center, Aarhus University Hospital. | Aarhus | 8200 | Denmark |
Not provided
| ID | Term |
|---|---|
| D010003 | Osteoarthritis |
| ID | Term |
|---|---|
| D001168 | Arthritis |
| D007592 | Joint Diseases |
| D009140 | Musculoskeletal Diseases |
| D012216 | Rheumatic Diseases |
Not provided
Not provided
Not provided
Not provided
Not provided
Not provided
| Up to 5 years |
| Osteolysis (radiolucent lines >1 mm) in the tibial periprosthetic region measured on radiographs | Up to 5 years |